  • - The study aimed to evaluate the accuracy and repeatability of the Iscan system in measuring force and pressure in the patellofemoral joint, crucial for understanding how mechanics relate to pain and cartilage degeneration.
  • - Results showed that the Iscan system had a root mean squared error of 6.5% for measuring resultant force and 0.86% for force distribution, with better accuracy for force distribution measurements.
  • - However, cementing the sensor to the cartilage negatively impacted the accuracy of force and pressure readings, indicating a trade-off between measurement method and sensor placement.

Article Abstract

The objective of this study was to assess how accurately and repeatably the Iscan system measures force and pressure in the natural patellofemoral joint. These measurements must be made to test widely held assumptions about the relationships between mechanics, pain and cartilage degeneration. We assessed the system's accuracy by using test rigs in a materials testing machine to apply known forces and force distributions across the sensor. The root mean squared error in measuring resultant force (for five trials at each of seven load levels) was 6.5 +/- 4.4% (mean +/- standard deviation over all trials at all load levels), while the absolute error was -5.5 +/- 5.6%. For force distribution, the root mean squared error (for five trials at each of five force distributions) was 0.86 +/- 0.58%, while the absolute error was -0.22 +/- 1.03%. We assessed the repeatability of the system's measurements of patellofemoral contact force, pressure and force distribution in four cadaver specimens loaded in continuous and static flexion. Variability in measurement (standard deviation expressed as a percentage of the mean) was 9.1% for resultant force measurements and 3.0% for force distribution measurements for static loads, and 7.3% for resultant force and 2.2% for force distribution measurements for continuous flexion. Cementing the sensor to the cartilage lowered readings of resultant force by 31 +/- 32% (mean +/- standard deviation), area by 24 +/- 13% and mean pressure by 9 +/- 34% (relative to the uncemented sensor). Maximum pressure measurement, however, was 24 +/- 43% higher in the cemented sensor than in the uncemented sensor. The results suggest that the sensor measures force distribution more accurately and repeatably than absolute force. A limitation of our work, however, is that the sensor must be cemented to the patellar articular surface to make the force distribution measurements, and our results suggest that this process reduces the accuracy of force, pressure and area measurements. Our results suggest that the Iscan system's pressure measurement accuracy and repeatability are comparable to that of Fuji Prescale film, but its advantages are that it is thinner than most Fuji Prescale film, it measures contact area more accurately and that it makes continuous measurements of force, pressure and area.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0021-9290(03)00105-2DOI Listing

Publication Analysis

Top Keywords

force distribution
24
force
17
force pressure
16
resultant force
16
pressure measurement
12
standard deviation
12
distribution measurements
12
+/-
10
accuracy repeatability
8
pressure
8

Similar Publications

Accurate models for predicting drop dynamics, such as maximum drop departure sizes, are crucial for estimating heat transfer rates during condensation on superhydrophobic (SH) surfaces. Previous studies have focused on examining the heat transfer rates for SH surfaces under the influence of gravity or vapor flowing over the surface. This study investigates the impact of surface solid fraction and texture scale on drop mobility in a condensing environment with a humid air flow.

View Article and Find Full Text PDF

Objective: The complex mix of factors, including hemodynamic forces and wall remodeling mechanisms, that drive intracranial aneurysm growth is unclear. This study focuses on the specific regions within aneurysm walls where growth occurs and their relationship to the prevalent hemodynamic conditions to reveal critical mechanisms leading to enlargement.

Methods: The authors examined hemodynamic models of 67 longitudinally followed aneurysms, identifying 88 growth regions.

View Article and Find Full Text PDF

Analysis of the Feasibility of the OrthoNail Hybrid Intramedullary Implant in the Human Body with Respect to Material Durability.

J Funct Biomater

January 2025

Department of Mechanics, Materials and Biomedical Engineering, Faculty of Mechanical Engineering, Wrocław University of Science and Technology, Smoluchowskiego 25, 50-372 Wrocław, Poland.

This study focuses on the development and evaluation of the OrthoNail hybrid intramedullary implant for lower limb lengthening in patients requiring significant skeletal reconstruction. The implant addresses the challenges in load-bearing during rehabilitation, providing a robust solution that is capable of supporting physiological loads. Mechanical tests, including axial compression, tension, torsion, and 3,4-point bending, determined the implant's load capacity and fatigue resistance, while finite element analysis assessed stress distributions in bone tissue and around screw holes during single-leg stance, with boundary conditions derived from Orthoload database data.

View Article and Find Full Text PDF

This study evaluated the potential usage of phosphorylated egg white protein (P-EWP) nanogels fabricated via microwave-induced phosphorylation modification and gel process and further ultrasonic nanometrization as novel delivery systems for cinnamon bark essential oil (CBEO). Compared to EWP-CBEO nanogels without chemical phosphorylation, the obtained P-EWP-CBEO nanogels have shown smaller average hydrodynamic diameter (133.6 nm), relatively uniform size distribution (polydispersity index around 0.

View Article and Find Full Text PDF

An Open-Source 3D Printed Three-Fingered Robotic Gripper for Adaptable and Effective Grasping.

Biomimetics (Basel)

January 2025

Department of Mechanical Engineering, Sungkyunkwan University, 2066 Seobu-ro, Jangan-gu, Suwon 16419, Republic of Korea.

This research focuses on the design of a three-finger adaptive gripper using additive manufacturing and electromechanical actuators, with the purpose of providing a low-cost, efficient, and reliable solution for easy integration with any robot arm for industrial and research purposes. During the development phase, 3D printing materials were employed in the gripper's design, with Polylactic Acid (PLA) filament used for the rigid mechanical components and Thermoplastic Polyurethane (TPU) for the flexible membranes that distribute pressure to the resistive force sensors. Stress analysis and simulations were conducted to evaluate the performance of the components under load and to gradually refine the design of the adaptive gripper.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!