Spectral shaping to improve the point spread function in optical coherence tomography.

Opt Lett

School of Optics, Center for Research and Education in Optics and Lasers, University of Central Florida, 4000 Central Florida Boulevard, Orlando, Florida 32816-2700, USA.

Published: October 2003

We demonstrate inhibition of the sidelobes of the axial point spread function in optical coherence tomography by shaping the power spectrum of the light source with a remaining power of 4.54 mW. A broadband amplified spontaneous emission source radiating at 1565 +/- 40 nm is employed in a free-space optical coherence tomography system. The axial point spread functions before and after optical spectral shaping are presented. Results show that spectral shaping of the source can inhibit sidelobes of the point spread function up to 12.9 dB, with an associated small increase of 2.2 dB in noise floor in the far field. The effect of spectral shaping on axial resolution is demonstrated according to three metrics. Image quality improvement is also illustrated with optical coherence tomography images of an onion before and after spectral shaping.

Download full-text PDF

Source
http://dx.doi.org/10.1364/ol.28.001921DOI Listing

Publication Analysis