Objective: Surgical intervention is an option for treating the remodeled and dilated left ventricles of patients with heart failure. Providing end-diastolic support with an innovative mesh-like cardiac support device reduces mechanical stress, improves function, and reverses cardiac remodeling in animal models without safety issues. The objective of this study was to review the global clinical safety and feasibility experience of this device.

Methods: The Acorn CorCap cardiac support device (Acorn Cardiovascular, Inc, St Paul, Minn) has been implanted worldwide in more than 130 patients with dilated cardiomyopathy with or without concomitant cardiac surgery. The device is positioned around the ventricles and given a custom fit. A series of 48 patients were implanted with the device in initial safety and feasibility studies, of whom 33 also received concomitant cardiac surgery.

Results: At implantation, 11 patients were in New York Heart Association class II, 33 were in class III, and 4 were in class IV. The average CorCap implantation time was 27 minutes. The mean intraoperative reduction in left ventricular end-diastolic dimension was 4.6% +/- 1%. There were no device-related intraoperative complications. Eight early and 9 late deaths occurred during follow-up extending to 24 months. Actuarial survival was 73% at 12 months and 68% at 24 months. There were no device-related adverse events or evidence of constrictive disease, and coronary artery flow reserve was maintained. Ventricular chamber dimensions decreased, whereas ejection fraction and New York Heart Association class were improved in patients overall and in those patients implanted with the CorCap device without concomitant operations.

Conclusions: The CorCap device appears safe for patients with dilated cardiomyopathy. Randomized clinical trials are underway in Europe, Australia, and North America.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0022-5223(03)00049-7DOI Listing

Publication Analysis

Top Keywords

cardiac support
12
support device
12
safety feasibility
8
patients dilated
8
dilated cardiomyopathy
8
concomitant cardiac
8
patients implanted
8
york heart
8
heart association
8
association class
8

Similar Publications

An atrial septal defect (ASD) is a common congenital heart anomaly that results in irregular blood flow between the systemic and pulmonary circulations due to an opening in the atrial septum. Ostium secondum ASD accounts for a large proportion of these defects and often goes unnoticed during childhood and adolescence. Pulmonary hypertension (PH), affecting a significant number of patients with ostium secondum ASD, is associated with functional limitations, heart failure, and tachyarrhythmias.

View Article and Find Full Text PDF

Infectious myocarditis (IM) and infective endocarditis (IE), sometimes associated with infection of the surrounding mediastinal tissue or embolic complications caused by residual implantable cardioverter defibrillator (ICD) lead material embedded in the ventricle, present a significant challenge for cardiac surgeons due to the difficulty of precisely locating the old intracardiac pacing lead remnants because of the heart's continuous movement. We present the case of successful two-stage elective sternotomy extraction of two residual defibrillator leads, one trapped in the left innominate vein, easily removed after veinotomy without cardiopulmonary bypass (CPB), and the other embedded intramyocardially in the inferior wall of the right ventricle, successfully removed under CPB after fluoroscopic guidance. The patient was discharged four weeks post-operation without complications.

View Article and Find Full Text PDF

Background: Within the United States, White individuals experience a higher risk of atrial fibrillation (AF) while exhibiting a lower AF-related stroke risk compared to other ethnic groups. It is possible that these observations stem from phenomena unique to the United States, such as differential health care access. The United Kingdom provides socialized medicine, which ostensibly promotes equitable health care access.

View Article and Find Full Text PDF

Background: The Hispanic/Latino population is not uniform. Prevalence and clinical outcomes of cardiac arrhythmias in ethnic background subgroups are variable, but the reasons for differences are unclear. Vectorcardiographic Global Electrical Heterogeneity (GEH) has been shown to be associated with adverse cardiovascular outcomes.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!