A simple and rapid reversed-phase high-performance liquid chromatographic method for the monitoring of process-related synthetic organic impurities of profenofos (PFS) is developed. Impurities are separated and determined on a reversed-phase Hypersil C(18) column using gradient elution of 50 mM ammonium formate buffer-acetonitrile as a mobile phase and detection at 230 nm at ambient temperature. The method is validated with respect to accuracy, precision, linearity, and limits of detection and quantitation. The method is found to be suitable not only for monitoring the reactions involved in the process development of PFS, but also quality assurance, as it can detect impurities at the level of 1.5 x 10(-8) g.

Download full-text PDF

Source
http://dx.doi.org/10.1093/chromsci/41.8.418DOI Listing

Publication Analysis

Top Keywords

liquid chromatographic
8
chromatographic method
8
method monitoring
8
monitoring process-related
8
process-related synthetic
8
synthetic organic
8
organic impurities
8
impurities profenofos
8
development validation
4
validation liquid
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!