Purpose: To investigate the repeat reliability of the multifocal visual evoked potential (mfVEP).

Patients And Methods: Fifteen subjects with no known abnormalities of the visual system and 10 patients with glaucoma participated in the study. Monocular mfVEPs were recorded on two separate days, using a 60-sector, pattern-reversal dart board array. Within a single session, two 7-minute. recordings were obtained for each eye. The amplitude of each mfVEP response was obtained using a root mean square measure (RMS). An mfVEP ratio [10*log (RMS day 1 / RMS day 2)] provided a measure of the reproducibility of an individual response. The same calculations were performed for Run 1 compared with Run 2 within a day and Run 1 (Run 2) compared with Run 1 (Run 2) across days.

Results: For all 1800 mfVEP responses (60 sectors x 15 subjects x 2 eyes), the correlation between the amplitude on day 2 and the amplitude on day 1 was good (r = 0.85). The mean standard deviation (SD) of the 60 mfVEP ratios for the individual subjects was 1.63 dB for the 14-minute records (the combination of the two 7-minute recordings). On average for the 7-minute records, the mean SD across days was 1.77 dB while the mean SD within a day was 1.53 dB. The correlation within a day (r = 0.87) also was slightly larger than across days (r = 0.80). The mean SD decreased as the RMS amplitude increased. The patients' mean SD was 1.75 dB with r equal to 0.82.

Conclusions: The repeat reliability of the mfVEP was good (approximately 1.6dB); in fact, it was better than that typically obtained with static automated perimetry (approximately 2.7dB). Repeat testing on separate days added surprisingly little to the variability seen with repeat testing within the same session.

Download full-text PDF

Source
http://dx.doi.org/10.1097/00061198-200310000-00002DOI Listing

Publication Analysis

Top Keywords

repeat reliability
12
reliability multifocal
8
multifocal visual
8
visual evoked
8
evoked potential
8
separate days
8
7-minute recordings
8
rms day
8
amplitude day
8
repeat testing
8

Similar Publications

Accurate diagnosis and monitoring of neurodegenerative diseases require reliable biomarkers. Cerebrospinal fluid (CSF) proteins are promising candidates for reflecting brain pathology; however, their diagnostic utility may be compromised by natural variability between individuals, weakening their association with disease. Here, we measured the levels of 69 pre-selected proteins in cerebrospinal fluid using antibody-based suspension bead array technology in a multi-disease cohort of 499 individuals with neurodegenerative disorders including Alzheimer's disease (AD), behavioral variant frontotemporal dementia, primary progressive aphasias, amyotrophic lateral sclerosis (ALS), corticobasal syndrome, primary supranuclear palsy, along with healthy controls.

View Article and Find Full Text PDF

Objective: Somatic variants causing epilepsy are challenging to detect, as they are only present in a subset of brain cells (e.g., mosaic), resulting in low variant allele frequencies.

View Article and Find Full Text PDF

Background: Alzheimer's Disease (AD) is a common neurodegenerative disorder affecting >35 million people worldwide. Despite extensive genetic studies, the identified factors only explain a small fraction of the heritable risk of AD. This suggests the contribution of yet-unknown genetic factors to the development of AD, such as tandem repeats (TRs).

View Article and Find Full Text PDF

A novel dual-mode detection method for microRNA-21 was developed. Photoluminescent (PL) and multiphonon resonant Raman scattering (MRRS) techniques were combined by using ZnTe nanoparticles as signal probes for reliable detection. The catalytic hairpin assembly (CHA) strategy was integrated with superparamagnetic FeO nanoparticle clusters (NCs) to enhance sensitivity.

View Article and Find Full Text PDF

There is limited knowledge about the size of the UK dog population. This makes it difficult to reliably monitor population dynamics and management. A repeatable method of measuring the UK dog population, including owned and unowned dogs i.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!