The purpose of this review was to evaluate the utility of the 20 MHz microvascular implantable Doppler probe for free-tissue transfer, both intra- and postoperatively. Over a 15-month period, the Doppler probe was used in a total of 260 anastomoses, including 118 arterial and 142 venous microanastamoses. In these 260 anastomoses, there were six false positive results and eight true positives, resulting in one flap loss. The free flap success rate was 99 percent, the re-exploration rate was 8 percent, and the salvage rate was 83 percent. The internal Doppler offers an easy and reliable way to monitor microvascular free-tissue transfer both intra- and postoperatively. This study demonstrates the continued increase in survival, as well as salvage, of free-tissue transfer.

Download full-text PDF

Source
http://dx.doi.org/10.1055/s-2003-42495DOI Listing

Publication Analysis

Top Keywords

doppler probe
12
free-tissue transfer
12
rate percent
12
internal doppler
8
transfer intra-
8
intra- postoperatively
8
260 anastomoses
8
retrospective review
4
review internal
4
doppler
4

Similar Publications

Identification of the Greater Palatine Foramen, the Greater Palatine Artery, and Palatal Fibromucosa: A Cross-Sectional Ultrasonographic Clinical Study.

J Stomatol Oral Maxillofac Surg

January 2025

Professor, Department of Oral, Dental and Maxillofacial Radiology, Faculty of Dentistry, Ataturk University, Erzurum, 25240, Turkiye.

Objectives: This study aimed to evaluate with ultrasonography (US) the location of GPF, the course of the GPA, and to measure palatal fibromucosa thickness.

Materials And Methods: This in vivo experiment was conducted on a group of healthy, young Turkish subjects. Using the US, the localization of the GPF, its diameter and opening type, maximum flow velocity (Vmax) and collateral branches (CB) of the GPA, and the thickness of the palatine mucosa were evaluated.

View Article and Find Full Text PDF

The nicotinic acetylcholine receptor (nAChR) is a pentameric ligand-gated ion channel (pLGIC) commonly used as a model for receptors belonging to the Cys-loop superfamily. Members of pLGICs are standardly used in numerous toxicological investigations e.g.

View Article and Find Full Text PDF

Introduction: Ultra-high-field magnetic resonance (MR) systems (7 T and 9.4 T) offer the ability to probe human brain metabolism with enhanced precision. Here, we present the preliminary findings from 3D MR spectroscopic imaging (MRSI) of the human brain conducted with the world's first 10.

View Article and Find Full Text PDF

Objective: To assess the agreement of transpulmonary thermodilution (TPTD) and transpulmonary ultrasound dilution (TPUD) against direct measurement of main pulmonary artery flow with an ultrasound transit time flow probe (UTF) over a wide range of conditions in anesthetized cats. Additionally, the trending ability of TPTD, TPUD, and esophageal Doppler ultrasonography (EDU) was evaluated against UTF.

Methods: 12 purpose-bred cats were used.

View Article and Find Full Text PDF

Recent studies have reported that monitoring spinal cord perfusion pressure (SCPP) using a pressure probe to measure "intraspinal pressure" (ISP) within the subdural space at the injury site may improve the hemodynamic management of acute spinal cord injury (SCI) patients. This study aimed to investigate, within a pig model of SCI, the relationship between the ISP measured within the subdural space and the "spinal cord pressure" (SCP) measured within the spinal cord itself. Specifically, we sought to characterize the changes to ISP and SCP over time, both rostral and caudal to the injury epicenter, and in relation to native spinal cord morphometry.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!