This study reports the dosimetry performed to support an experiment that measured physiological responses of volunteer human subjects exposed to the resonant frequency for a seated human adult at 100 MHz. Exposures were performed in an anechoic chamber which was designed to provide uniform fields for frequencies of 100 MHz or greater. A half wave dipole with a 90 degrees reflector was used to optimize the field at the subject location. The dosimetry plan required measurement of transmitter harmonics, stationary probe drift, field strengths as a function of distance, electric and magnetic field maps at 200, 225, and 250 cm from the dipole antenna, and specific absorption rate (SAR) measurements using a human phantom, as well as theoretical predictions of SAR with the finite difference time domain (FDTD) method. On each exposure test day, a measurement was taken at 225 cm on the beam centerline with a NBS E field probe to assure consistently precise exposures. A NBS 10 cm loop antenna was positioned 150 cm to the right, 100 cm above, and 60 cm behind the subject and was read at 5 min intervals during all RF exposures. These dosimetry measurements assured accurate and consistent exposures. FDTD calculations were used to determine SAR distribution in a seated human subject. This study reports the necessary dosimetry for work on physiological consequences of human volunteer exposures to 100 MHz.