Purpose: To evaluate stent-graft and aneurysm wall motions during the cardiac cycle using cine magnetic resonance imaging (MRI) to identify mechanisms of long-term failure of endovascular aneurysm repair (EVAR).

Methods: Prior to and after EVAR in 7 patients with abdominal aortic aneurysms (AAA), 12 MRI images per cardiac cycle were acquired in transverse, sagittal, and coronal planes of the aneurysm. Two independent observers blinded to the aim of the study manually traced stent-graft and aneurysm wall contours. Translation was defined as the maximal displacement of the contours in the peak-systolic image compared to the end-diastolic image. Aneurysm wall motions before and after repair were compared. Stent-graft and aneurysm configuration changes during the cardiac cycle were evaluated. The relation between translation and the degree of angulation of the stent-graft was calculated.

Results: The anteroposterior translation of the aneurysm decreased from a median 1.05 mm (range <0.5-1.29) before EVAR to within pixel size (<0.5 mm) after EVAR (p=0.04). The cranial-caudal translation of the aneurysm increased from a median 1.01 mm (range <0.5-1.51) before to 1.69 mm (range 1.1-1.99) after EVAR (p=0.02). In 4 stent-grafts, bending during cardiac systole was observed at the site of maximal angulation of the device. In transverse sections, 2-dimensional pulsatile wall motion of the aneurysm was 0.25 cm(2) (range 0.07-0.29) before and 0.17 cm(2) (range 0.07-0.42) after EVAR (p=0.79). No pulsatility of the stent-graft itself was observed. The correlation coefficient between angulation of the stent-graft and the increase in cranial-caudal translation after EVAR was 0.67 (p>0.05).

Conclusions: After EVAR, increased longitudinal translation of both the aneurysm and stent-graft was observed, indicating downward pulling forces at the proximal fixation site. Secondly, increased bending was seen at the site of maximal angulation, which implies a risk of metal fatigue and fabric damage at sites of stent-graft angulation.

Download full-text PDF

Source
http://dx.doi.org/10.1177/152660280301000306DOI Listing

Publication Analysis

Top Keywords

stent-graft aneurysm
12
aneurysm wall
12
cardiac cycle
12
aneurysm
8
wall motions
8
cine mri
4
mri assessment
4
assessment aortic
4
aortic aneurysm
4
aneurysm dynamics
4

Similar Publications

Objective: We present our experience with endovascular Bentall procedure (Endo-Bentall) using a modular valve conduit (Endo-Bentall) in high-risk patients with aortic root pathologies.

Methods: The physician constructed Endo-Bentall device is composed of a self-expanding transcatheter aortic valve (TAVR), aortic endovascular stent graft (TEVAR), and two wire-reinforced fenestrations for coronary artery stenting. The TAVR valve is sutured into an appropriately sized TEVAR graft.

View Article and Find Full Text PDF

Objective: Half of re-interventions after fenestrated and branched endovascular aortic repair (FB-EVAR) are target vessel related. Regarding bridging stent choice, existing data are controversial. This meta-analysis aimed to evaluate the performance of Advanta V12/iCAST as bridging stent in FB-EVAR.

View Article and Find Full Text PDF

Endoanchors for the distal fixation of iliac limb in endovascular aneurysm repair.

J Vasc Surg Cases Innov Tech

April 2025

Division of Vascular and Endovascular Surgery, Department of Cardiac, Thoracic, Vascular Sciences and Public Health, University of Padua, Padua, Italy.

This report details the case of an 84-year-old male with an infrarenal abdominal aortic aneurysm and a dilated right common iliac artery eligible for endovascular treatment. A bifurcated stent graft (Medtronic Endurant IIs) was used to treat the aneurysm. To address the concerns of instability of the right iliac limb, four endoanchors (Heli-FX EndoAnchor, Medtronic) were placed at the distal landing zone to provide additional fixation.

View Article and Find Full Text PDF

Stanford type B aortic dissection involving the left subclavian artery (LSA) poses significant clinical challenges. The Castor single-branch stent graft and in situ fenestration are commonly used techniques, but the better endovascular treatment remains debated. This study evaluates the clinical effects of the Castor single-branched stent graft versus in situ fenestration in treating Stanford type B aortic dissection involving the LSA.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!