Unlabelled: Heart to mediastinum count ratio (H/M) has been commonly utilized as an indicator of myocardial I-123 MIBG uptake. However, normal ranges of H/M were markedly different among various gamma camera systems. The purpose of this study was to clarify whether scatter correction by two-window method standardizes H/M among various gamma camera systems.

Methods: Scatter uncorrected and corrected MIBG imaging was acquired in phantom and human studies in combination with low energy high-resolution collimator (LEHR) and medium energy collimator (MEC). For scatter correction, energy window width of 159 keV +/- 10% was applied to main window imaging and 193 keV +/- 9.5% was applied to upper window imaging for scatter correction.

Results: In phantom study, a significant difference was observed in uncorrected H/M among three gamma camera systems using LEHR or MEC (2.09 +/- 0.06 vs. 2.58 +/- 0.03 in GCA7200 camera, 2.00 +/- 0.07 vs. 2.42 +/- 0.06 in DS7 camera and 2.16 +/- 0.04 vs. 2.67 +/- 0.07 in Vertex plus camera). However, there was no significant difference in corrected H/M among the three gamma camera systems, either with LEHR or MEC (2.70 +/- 0.07 vs. 2.69 +/- 0.07 in GCA7200 camera, 2.66 +/- 0.08 vs. 2.61 +/- 0.05 in DS7 camera and 2.66 +/- 0.05 vs. 2.61 +/- 0.05 in Vertex plus camera). In human study, uncorrected H/M in DS7 camera with LEHC was significantly lower than that in GCA7200 camera with MEC (1.60 +/- 0.37 vs. 1.85 +/- 0.54, N = 14). In contrast, the difference was insignificant in corrected H/M (2.12 +/- 0.59 vs. 2.16 +/- 0.68). There was a very excellent correlation in corrected H/M between DS7 and GCA7200 cameras (r = 0.991, p < 0.001).

Conclusion: This study demonstrated that scatter correction by the two-window method standardizes the H/M in MIBG scintigraphy either with LEHR or MEC. Scatter corrected H/M can be applied to measure a standardized parameter of MIBG uptake in human clinical studies using various gamma camera systems.

Download full-text PDF

Source
http://dx.doi.org/10.1007/BF02988527DOI Listing

Publication Analysis

Top Keywords

gamma camera
24
camera systems
20
+/-
18
scatter correction
16
+/- 007
16
corrected h/m
16
camera
14
correction two-window
12
two-window method
12
method standardizes
12

Similar Publications

CoReSi: a GPU-based software for Compton camera reconstruction and simulation in collimator-free SPECT.

Phys Med Biol

January 2025

CREATIS, INSA de Lyon, Bâtiment Blaise Pascal, 7 Avenue Jean Capelle, Villeurbanne, 69621 Cedex , FRANCE.

Compton cameras are imaging devices that may improve observation of sources of γ photons. We present CoReSi, a Compton Reconstruction and Simulation software implemented in Python and powered by PyTorch to leverage multi-threading and for easy interfacing with image processing and deep learning algorithms. The code is mainly dedicated to medical imaging and for near-field experiments where the images are reconstructed in 3D.

View Article and Find Full Text PDF

Aims: While most clinical guidelines recommend using a 64-projection view technique, some protocols do not specify a preference between 32-projection and 64-projection methods for conducting myocardial perfusion scintigraphy (MPS), which shows the lack of consensus in this matter. Nevertheless, these guidelines and protocols have not provided us with compelling evidence to support why the 64-projection technique is usually chosen. Thus, we aimed to determine if there is a significant difference between them in the assessment of cardiac perfusion and functional indices.

View Article and Find Full Text PDF

This paper explores the adaptation and application of i-TED Compton imagers for real-time dosimetry in Boron Neutron Capture Therapy (BNCT). The i-TED array, previously utilized in nuclear astrophysics experiments at CERN, is being optimized for detecting and imaging 478 keV gamma-rays, critical for accurate BNCT dosimetry. Detailed Monte Carlo simulations were used to optimize the i-TED detector configuration and enhance its performance in the challenging radiation environment typical of BNCT.

View Article and Find Full Text PDF

GFR Estimation and Correlation for Oncology Patients by Two Methods, Gates Method and Dual Time Point Plasma Sampling Method.

Indian J Nucl Med

November 2024

Department of Nuclear Medicine and Molecular Imaging, Homi Bhabha Cancer Hospital and Mahamana Pandit Madan Mohan Malaviya Cancer Centre, Tata Memorial Centre, Homi Bhabha National Institute (HBNI), Varanasi, India.

Background: With the increasing number of oncology cases and a parallel surge in chemotherapeutic drugs for treatment, the treating physicians conducts nephrotoxicity evaluation to provide a personalized dosing strategy. Of the various tests available, glomerular filtration rate (GFR) under gamma camera with help of Gates method has gained importance, being a good index of overall kidney functions. In addition to this, there has been an alternate and old method for GFR estimation: plasma sampling.

View Article and Find Full Text PDF

Background: High-dose-rate (HDR) brachytherapy using Iridium-192 as a radiation source is widely employed in cancer treatment to deliver concentrated radiation doses while minimizing normal tissue exposure. In this treatment, the precision with which the sealed radioisotope source is delivered significantly impacts clinical outcomes.

Purpose: This study aims to evaluate the feasibility of a new four-dimensional (4D) in vivo source tracking and treatment verification system for HDR brachytherapy using a patient-specific approach.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!