Purpose: To compare scleral Tono-Pen (Medtronic Solan, Jacksonville, FL) readings to corneal Tono-Pen readings.

Methods: Intraocular pressure (IOP) was measured prospectively in 72 eyes of 37 adult patients and in 10 eyes of 5 children. Measurements were taken on the central cornea and on the sclera. Recorded measurements were within 5% confidence levels by Tono-Pen.

Results: Corneal IOP ranged from 10 to 28 mm Hg (mean +/- standard deviation, 17.0 +/- 3.8 mm Hg). Scleral measurements ranged from 4 to 99 mm Hg (40.4 +/- 23.0 mm Hg). Scleral measurements ranged from 11 mm Hg lower to 76 mm Hg higher than corneal measurements.

Conclusions: Tono-Pen readings obtained from sclera are of no clinical value and should not be used to approximate corneal IOP.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s1091-8531(03)00147-2DOI Listing

Publication Analysis

Top Keywords

corneal tono-pen
8
tono-pen readings
8
corneal iop
8
scleral measurements
8
measurements ranged
8
corneal
5
comparison scleral
4
scleral corneal
4
tono-pen
4
readings purpose
4

Similar Publications

Background: To evaluate the agreement between the Goldman applanation tonometer (GAT), Tono-Pen, and noncontact tonometer (NCT) in the measurement of intraocular pressure (IOP) in pseudophakic children.

Methods: The medical records of nonglaucomatous pseudophakic children between 2009 and 2019 were retrospectively analyzed. A total of 46 eyes of 23 patients operated for bilateral pediatric cataract were included in the study.

View Article and Find Full Text PDF

Purpose: The aim of this study was to evaluate the effects of different stages of diabetic retinopathy (DR) and metabolic control of blood glucose levels on corneal biomechanical parameters.

Methods: Diabetic patients were categorized into three groups: no DR group, nonproliferative DR (NPDR) group, and proliferative DR (PDR) group. Of the 141 eyes examined, 40 belonged to the control group, 34 to no DR group, 34 to NPDR group, and 33 to PDR group.

View Article and Find Full Text PDF
Article Synopsis
  • This review discusses various tonometers used to measure intraocular pressure (IOP) in children with glaucoma, focusing on their accuracy, tolerability, and ease of use.
  • Goldmann applanation tonometry (GAT) is considered the gold standard, while Tono-Pen and rebound tonometer (RBT) are also effective but can overestimate IOP, particularly in cases of high pressure or corneal issues.
  • RBT is recommended for pediatric use due to its ease of use and lower traumatic impact, but it's advised to confirm high IOP readings with a second tonometer.
View Article and Find Full Text PDF
Article Synopsis
  • * Results showed that measurements from the different tonometers were comparable, with strong correlations and good agreement, particularly between the Icare and Tono-Pen.
  • * The findings suggest that these alternative IOP measurement devices are reliable, potentially enhancing clinical practices in diagnosing and managing eye conditions.
View Article and Find Full Text PDF

Purpose: To compare intraocular pressure (IOP) obtained with Tono-Pen (TP) and Goldmann applanation (GAT) using large-scale electronic health records (EHR).

Design: Retrospective cohort study.

Methods: A single pair of eligible TP/GAT IOP readings was randomly selected from the EHR for each ophthalmology patient at an academic ophthalmology center (2013-2022), yielding 4550 eligible measurements.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!