Background: The prosthesis used for aortic valve replacement (AVR) can be too small in relation to body size, thus causing valve prosthesis-patient mismatch (PPM) and abnormally high transvalvular pressure gradients. This study examined if there is a relation between PPM and short-term mortality after operation.
Methods And Results: The indexed valve effective orifice area (EOA) was estimated for each type and size of prosthesis being implanted in 1266 consecutive patients and used to define PPM as not clinically significant if >0.85 cm2/m2, as moderate if >0.65 cm2/m2 and
Conclusions: PPM is a strong and independent predictor of short-term mortality among patients undergoing AVR, and its impact is related both to its degree of severity and the status of left ventricular function. In contrast to other risk factors, moderate-severe PPM can be largely avoided with the use of a prospective strategy at the time of operation.