Comparison of two microenvironments and nurse caregiving on thermal stability of ELBW infants.

Adv Neonatal Care

College of Nursing and NICU, University of Utah, 10 South 2000 East Front, Salt Lake City, UT 84112, USA.

Published: June 2002

AI Article Synopsis

  • The study evaluated a new bioinstrumented radiant warmer (RINCE) to see if it could help stabilize body and ambient temperatures in extremely low birth-weight infants during caregiver interruptions.
  • The sample included 10 infants with an average birth weight of 730 grams, and the research used a crossover time-series design to compare two different bed conditions.
  • Results showed that RINCE maintained higher heel temperatures and reduced temperature differences compared to the control bed, although overall abdominal temperature stability was similar in both environments.

Article Abstract

Purpose: This study tested a novel bioinstrumented radiant warmer (RW), designed to provide a reduced-intervention nursing caregiving environment (RINCE) on the ambient and body surface temperature stability of extremely low birth-weight (ELBW) infants (< 1,000 g) during caregiver disruptions. The RINCE was compared with a control bed consisting of a standard RW, modified with a Plexiglas hood (Rohm and Haas, Philadelphia, PA) and an external humidity source.

Subjects: The sample consisted of 10 extremely low birth-weight infants. The mean gestational age was 25.7 weeks (+/- 1.4 weeks). Birth weight ranged from 570 to 880 g: mean birth weight was 730 g (+/- 95.14 g). The mean Scores for Neonatal Acute Physiology was 21.4 (+/- 6).

Design: This clinical study used a small n sample, crossover time-series design. Infants were randomized to treatment order (control bed v RINCE) on the first day of life. A washout period between microenvironments was observed, and then each infant was placed into the other condition.

Methods: Each infant was studied for 5 hours in each treatment condition (control bed v RINCE). Abdominal skin temperatures, heel temperatures, ambient temperatures, and relative humidity (RH) were recorded by a computerized data acquisition system every 60 seconds for hours 12 to 24 of life. Caregiver interventions, microenvironment disruptions, or both were manually coded in the data set.

Principal Results: The mean number of minutes abdominal temperatures were less than 36.5 degrees C was not significantly different in either microenvironment (P = 0.48). The mean number of minutes that heel temperatures were less than 35.2 degrees C and central-peripheral differences were greater than 2 degrees C was significantly less in the RINCE (P < 0.04 and P < 0.0001, respectively). Although the number and duration of caregiving disruptions (P = 0.71), number of procedures (P = 0.30), and the duration of interventions (P = 0.56) were not significantly different in either environment, the ambient air temperature (P < 0.001) and RH (P < 0.001) were significantly less variable in the RINCE during and after caregiving disruptions.

Conclusions: This study identified the extreme inter-dependence of the ELBW infant's temperature and the ambient environment. The RINCE significantly improved ambient temperature variability, RH variability, and the infant's peripheral and delta temperature (delta T) stability, particularly during and after caregiver disruptions.

Download full-text PDF

Source
http://dx.doi.org/10.1053/adnc.2002.33543DOI Listing

Publication Analysis

Top Keywords

control bed
12
elbw infants
8
environment rince
8
extremely low
8
low birth-weight
8
caregiver disruptions
8
birth weight
8
bed rince
8
heel temperatures
8
number minutes
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!