Neuronavigation accuracy dependence on CT and MR imaging parameters: a phantom-based study.

Phys Med Biol

Department of Clinical Physiopathology, Medical Physics Unit, University of Florence, Florence, Italy.

Published: July 2003

Clinical benefits from neuronavigation are well established. However, the complexity of its technical environment requires a careful evaluation of different types of errors. In this work, a detailed phantom study which investigates the accuracy in a neuronavigation procedure is presented. The dependence on many different imaging parameters, such as field of view, slice thickness and different kind of sequences (sequential and spiral for CT, T1-weighted and T2-weighted for MRI), is quantified. Moreover, data based on CT images are compared to those based on MR images, taking into account MRI distortion. Finally, the contributions to global accuracy coming from image acquisition, registration and navigation itself are discussed. Results demonstrate the importance of imaging accuracy. Procedures based on CT proved to be more accurate than procedures based on MRI. In the former, values from 2 to 2.5 mm are obtained for 95% fractiles of cumulative distribution of Euclidean distances between the intended target and the reached one while, in the latter, the measured values range from 3 to 4 mm. The absence of imaging distortion proved to be crucial for registration accuracy in MR-based procedures.

Download full-text PDF

Source
http://dx.doi.org/10.1088/0031-9155/48/14/311DOI Listing

Publication Analysis

Top Keywords

dependence imaging
8
imaging parameters
8
based images
8
procedures based
8
neuronavigation accuracy
4
accuracy dependence
4
imaging
4
parameters phantom-based
4
phantom-based study
4
study clinical
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!