Enzymatic determination of pyrophosphate in urine by flow methods.

Anal Sci

Department of Chemistry, University of Balearic Islands, E 07071 Palma de Mallorca, Spain.

Published: July 2003

Two flow methods for the enzymatic determination of pyrophosphate are described that are used to diminish the consumption of reagents. One method is based on the use of an open-close circuit with manual injection using a syringe. The other is a sequential injection method. The analytical features of both methods are: a linear range of 0.4 - 20 mg L(-1), an LOD of 0.38 mg L(-1), and a CV of 2.0% for the sequential injection method, and a linear range of 0.3 - 15 mg L(-1), an LOD of 0.29 mg L(-1), and CV of 2.2% for the open-close circuit method. The methods were applied to the determination of pyrophosphate in urine. The pyrophosphate concentration determined in urine samples varied from 1.26 to 6.67 mg L(-1).

Download full-text PDF

Source
http://dx.doi.org/10.2116/analsci.19.1029DOI Listing

Publication Analysis

Top Keywords

determination pyrophosphate
12
enzymatic determination
8
pyrophosphate urine
8
flow methods
8
open-close circuit
8
sequential injection
8
injection method
8
linear range
8
range l-1
8
l-1 lod
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!