Background And Objectives: The Hancock II bioprosthesis is a second-generation porcine valve xenograft treated with the detergent sodium dodecyl sulphate (T6) to retard calcification. The aim of this investigation was to study the gross and microscopic features in Hancock II explants to assess the structural changes occurring with time.

Methods: Among 1382 Hancock II bioprostheses (701 isolated aortic, 421 isolated mitral, 130 double) implanted from 1983 to 1997 in 1252 patients, 22 (16 mitral, 6 aortic) were removed at reoperation until 1999 and were available for pathological investigation: infective endocarditis occurred in 5 and structural deterioration in 8, whereas in the remaining 9 xenografts reoperation was performed for nonstructural valve deterioration (paravalvular leak in 4 and prophylactic replacement in 5). Morphological investigation consisted of gross examination and x-ray, histologic, immunohistochemistry, electron microscopic, and atomic absorption spectroscopic examination.

Results: The cause of structural valve deterioration was dystrophic calcification in 4 cases (1 aortic, 3 mitral; range of time graft was in place, 101 to 144 months), non-calcium-related tears in 3 cases (all mitral, range 121 to 163 months), and commissural dehiscence in 1 (aortic, range 156 months). Five of the nonstructural valve deterioration explants (range 42 to 122 months) showed only pinpoint mineralization at the commissures. Mean calcium content in nonstructural deterioration explants was 14.70 +/- 22.33 versus 99.11 +/- 81.52 mg/g in explants with structural valve deterioration. Electron microscopic examination showed early nuclei of mineralization mostly consisting of calcospherulae upon cell debris. Local or diffuse lipid insudation was observed in all but 2 explants and consisted of cholesterol clefts, lipid droplets, and lipid-laden macrophages featuring foam cells. The lipid insudation was the most plausible cause of tearing in 2 explants.

Conclusions: These pathologic findings support the clinical results of a delayed occurrence of structural failure of Hancock II bioprostheses and a mitigation of mineralization by the anti-calcification treatment. However, other factors such as lipid insudation may come into play in the long term.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0022-5223(03)00131-4DOI Listing

Publication Analysis

Top Keywords

valve deterioration
16
lipid insudation
12
hancock bioprosthesis
8
hancock bioprostheses
8
nonstructural valve
8
electron microscopic
8
structural valve
8
mitral range
8
deterioration explants
8
explants
6

Similar Publications

Hemodynamic Valve Deterioration After Transcatheter Aortic Valve Replacement: Incidence, Predictors, and Clinical Outcomes.

JACC Cardiovasc Interv

January 2025

Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland. Electronic address:

Background: Reports on the durability of transcatheter aortic valve replacement (TAVR) prostheses are scarce and confounded by varying definitions and competing risks of death.

Objectives: The authors sought to determine the incidence, predictors, and clinical outcomes of hemodynamic valve deterioration (HVD) according to the Valve Academic Research Consortium 3 definition after TAVR.

Methods: We analyzed consecutive patients undergoing TAVR in the prospective Bern TAVI (Transcatheter Aortic Valve Implantation) registry between August 2007 and June 2022 for the incidence and predictors of HVD and performed case control-matching to compare outcomes according to HVD.

View Article and Find Full Text PDF

Background: Lifetime treatment of aortic valve disease is a matter of increasing debate. Although the risks of a second aortic valve intervention are recognized, little attention has been given to the challenges of a third.

Objectives: This study delves into the clinical characteristics, indications, and outcomes of patients undergoing 3 aortic valve interventions.

View Article and Find Full Text PDF

Objective: To describe the characteristics of patients diagnosed with acute heart failure (AHF) in emergency departments (EDs) who develop cardiogenic shock (CS) not associated with ST-segment elevation acute coronary syndrome (STACS).

Methods: Information for patients diagnosed with AHF in 23 Spanish EDs and registered between 2009 and 2019 were included for analysis if the patients developed symptoms consistent with CS. We described baseline clinical characteristics related to cardiac decompensation and CS, as well as 30-day mortality.

View Article and Find Full Text PDF

Background: Transcatheter aortic valve implantation (TAVI) is a safe and effective therapy for patients with severe aortic stenosis. A Stuck leaflet and severe intraprosthetic regurgitation after valve implantation occur rarely but can lead to sudden haemodynamic deterioration. We encountered a case of a stuck leaflet following post-dilatation with the Edwards Sapien 3 Ultra RESILIA valve.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!