Study Objectives: In light of the ongoing debate over the clinical significance of periodic limb movement disorder (PLMD) and its monitoring in overnight sleep studies, we introduce a novel, portable, low-cost device for PLMD testing. The KickStrip is a disposable device, which includes a movement sensor, a central processing unit with real-time software, and a display. In the present study, the KickStrip final score (Kscore) is validated against the traditional periodic limb movement index (PLMI) based on overnight recordings in the sleep laboratory. DEISGN: Patients underwent full polysomnographic recordings concomitantly with the use of the KickStrip for a single night.
Setting: Sleep Disorders Unit at Loewenstein Rehabilitation Hospital and Sleep Medicine Center at Rambam Medical Center, Israel.
Patients: Eighty-two patients referred for overnight sleep recordings due to sleep disturbance of any kind.
Interventions: N/A.
Measurements And Results: The Kscores were collected and the PLMI computed for each leg separately. Pearson correlations between Kscores and PLMI ranged between r = 0.83 to r = 0.88 (p < 0.001). Sensitivity and specificity values of the Kscores for increasing PLMD thresholds showed sensitivity ranging from 50% to 100% and specificity ranging from 83% to 100%. Receiver operating characteristic curves showed area-under-the-curve values ranging from 82% to 92%. Bland-Altman plot showed high agreement between the methods.
Conclusions: Comparisons between the Kscores and the traditional PLMI show increased accuracy with severity level and excellent agreement. The KickStrip is a valuable tool for PLMD testing for both clinical and research purposes.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1093/sleep/26.4.480 | DOI Listing |