Background: Activated clotting time (ACT) is used to monitor heparin therapy during cardiopulmonary bypass, interventional cardiology, and hemodialysis. Traditionally, ACT is performed by use of the Hemochron system. Recently, a new device, the i-STAT system, has been introduced to measure ACT. The aim of this study was to correlate the performances of these two systems and to compare ACT values with heparin levels.
Methods: One hundred sixty-five samples from 29 patients undergoing cardiopulmonary bypass or hemodialysis were assayed in duplicate with two Hemochron and two i-STAT devices. Heparin levels were determined by anti-factor Xa assay.
Results: The Hemochron ACT ranged from 88 to 1,028 s, and the i-STAT ACT ranged from 80 to 786 s. Heparin plasma levels ranged from 0.01 to 10.8 U/mL. Bland-Altman analysis showed a mean difference between the two methods of 24 +/- 101 s. Strong relationships between anti-factor Xa activity and Hemochron ACTs (r2 = 0.69, P < 0.001) and i-STAT ACTs (r2 = 0.79, P < 0.001) were observed. During cardiac surgery, significant correlations were found: Hemochron, r2 = 0.61, P < 0.001 and i-STAT, r2 = 0.74, P < 0.001. During hemodialysis, relationships between anti-factor Xa activity and ACTs were found: Hemochron, r2 = 0.62, P < 0.001 and i-STAT, r2 = 0.55, P < 0.001.
Conclusions: During cardiopulmonary bypass procedure and hemodialysis, i-STAT provides measurements of clotting time quite similar to Hemochron ACT, which were significantly correlated with heparin levels.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/00000542-200307000-00012 | DOI Listing |
J Soc Cardiovasc Angiogr Interv
December 2024
Division of Cardiovascular Medicine, Sulpizio Cardiovascular Center, University of California San Diego, San Diego, California.
Background: A minimum threshold activated clotting time (ACT) to guide heparin dosing during percutaneous coronary intervention (PCI) is associated with lower ischemic complications. However, data are variable regarding the risk of high ACT levels. The aim of this study was to assess the impact of peak procedural ACT on complications and mortality for transfemoral and transradial access PCI.
View Article and Find Full Text PDFClin Exp Emerg Med
January 2025
Department of Pharmacy Practice, Faculty of Pharmacy, King Abdulaziz University, Jeddah, Saudi Arabia.
Objectives: The objective of this systematic review and meta-analysis is to evaluate the efficacy, safety, time to INR reversal, total volume of 4-factor prothrombin complex concentrate (PCC) administered of fixed-dose versus weight-based dosing strategies in patients requiring urgent warfarin reversal, with specific focus on clinical outcomes such as hemostatic efficacy, thromboembolic events, and mortality rates.
Methods: A comprehensive systematic review was conducted using the PubMed, Embase, and Cochrane databases from inception through October 2023. We searched for randomized clinical trials or observational studies that compared efficacy or safety outcomes of fixed-dose vs.
Math Biosci Eng
December 2024
Laboratory of Optimization, Design, and Advanced Control, School of Chemical Engineering, Universidade Estadual de Campinas, Campinas, São Paulo, Brazil.
In the pursuit of personalized medicine, there is a growing demand for computational models with parameters that are easily obtainable to accelerate the development of potential solutions. Blood tests, owing to their affordability, accessibility, and routine use in healthcare, offer valuable biomarkers for assessing hemostatic balance in thrombotic and bleeding disorders. Incorporating these biomarkers into computational models of blood coagulation is crucial for creating patient-specific models, which allow for the analysis of the influence of these biomarkers on clot formation.
View Article and Find Full Text PDFZhonghua Yi Xue Za Zhi
January 2025
Shanghai Key Laboratory of Kidney and Blood Purification, Shanghai Medical Center of Kidney, Shanghai200032, China.
To investigate anticoagulation effects of nafamostat mesylate(NM) in sustained low-efficiency dialysis (SLED) and its relevant factors. Critically ill patients with kidney disease who were admitted to Zhongshan Hospital Affiliated to Fudan University and underwent SLED treatment from May to August 2024 were retrospectively included. Baseline clinical data were collected, and the activated partial thromboplastin time (APTT) and activated clotting time (ACT) were measured at the arterial end, before the filter, and at the venous end two hours post-NM anticoagulation treatment.
View Article and Find Full Text PDFMicrobiol Spectr
January 2025
Department of Biological Sciences, University of Notre Dame, Notre Dame, Indiana, USA.
Unlabelled: Group A (GAS) is a major human pathogen that causes several invasive diseases including necrotizing fasciitis. The host coagulation cascade initiates fibrin clots to sequester bacteria to prevent dissemination into deeper tissues. GAS, especially skin-tropic bacterial strains, utilize specific virulence factors, plasminogen binding M-protein (PAM) and streptokinase (SK), to manipulate hemostasis and activate plasminogen to cause fibrinolysis and fibrin clot escape.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!