Severity: Warning
Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Rationale And Objectives: The purpose of this study was (a) to demonstrate the feasibility of computing T1rho maps of, and T1rho dispersion in, human wrist cartilage at MR imaging in vivo and (b) to compare T1rho and T2 weighting in terms of magnitude of relaxation times and signal intensity contrast.
Materials And Methods: T2 and T1rho magnetic resonance images of wrist joints in healthy volunteers (n = 5) were obtained with a spin-echo sequence and a fast spin-echo sequence pre-encoded with a spin-lock pulse cluster. A 1.5-T clinical imager was used (Signa; GE Medical Systems, Milwaukee, Wis) with a 9.5-cm-diameter transmit-receive quadrature birdcage coil tuned to 63.75 MHz.
Results: T1rho relaxation times at a spin-lock frequency of 500 Hz vary from 40.5 msec +/- 0.85 to 56.6 msec +/- 4.83, and T2 relaxation times vary from 28.1 msec +/- 1.88 to 34.5 msec +/- 2.63 (mean +/- standard error of the mean, n = 5, P < .016) in various regions of the wrist. T1rho dispersion was observed in the range of spin-lock frequencies studied. T1rho-weighted images not only have higher signal-to-noise ratios but also show better fluid and fat signal suppression than T2-weighted images.
Conclusion: It was possible to perform T2- and T1rho-weighted MR imaging of human wrist cartilage in vivo with standard clinical imagers. The higher signal-to-noise ratio and improved contrast between cartilage and surrounding fat achieved with T1rho imaging may provide better definition of lesions and accurate quantitation of small changes in cartilage degeneration.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/s1076-6332(03)80079-x | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!