Background And Objective: When the endotracheal tube cuff is repeatedly aspirated to avoid excessive cuff pressure during nitrous oxide anaesthesia, a stable cuff pressure is eventually achieved. We assessed the time required to achieve a stable cuff pressure after repeated cuff deflation.

Methods: During 67% nitrous oxide and oxygen anaesthesia, air-filled cuffs of a standard tracheal tube (Mallinckrodt Hi-Contour) were repeatedly deflated every 30 min for the first 3 or 4 h to inhibit excessive pressure (Groups Def-3 or Def-4, respectively, n = 10 for each); the cuff pressure was monitored for an additional 3 h. In some patients, the study was terminated at 1, 2, 3 and 4 h (n = 6 for each).

Results: Cuff pressure in Group Def-3, but not in Group Def-4, > 22 mmHg after stopping cuff aspiration. Intracuff nitrous oxide concentrations increased during repeated cuff deflation and increased further in Group Def-3 during an additional 3 h (from 39.8 +/- 4.7% to 44.3 +/- 3.8%; P < 0.05), whereas intracuff nitrous oxide concentrations at 4 h were not different from those in Group Def-4 at the end of the study (43.7 +/- 4.5% versus 42.3 +/- 4.8%; P = 0.579).

Conclusions: When the air-filled cuff of the standard endotracheal tube is repeatedly deflated every 30 min for 4 h, but not for only 3 h, during nitrous oxide anaesthesia, a stable cuff pressure can be achieved without further deflation of the cuff. Our data also suggest that achieving an equilibrating nitrous oxide concentration in the cuff provides a subsequent stable cuff pressure.

Download full-text PDF

Source
http://dx.doi.org/10.1017/s0265021503000747DOI Listing

Publication Analysis

Top Keywords

cuff pressure
32
nitrous oxide
28
stable cuff
20
cuff
16
pressure
9
time required
8
required achieve
8
achieve stable
8
pressure repeated
8
endotracheal tube
8

Similar Publications

Pocket hematoma is a common and serious complication following cardiac implantable electronic device (CIED) implantation, contributing to significant morbidity and mortality. This study aimed to evaluate the efficacy of a novel pocket compression device in reducing pocket hematoma occurrence. We enrolled 242 patients undergoing CIED implantation, randomly assigning them to receive either the novel compression vest with a pressure cuff or conventional sandbag compression.

View Article and Find Full Text PDF

Noise pollution is a known health risk factor and evidence for cardiovascular diseases associated with traffic noise is growing. At least 20% of the European Union's population lives in noise-polluted areas with exposure levels exceeding the recommended limits of the World Health Organization, which is considered unhealthy by the European Environment Agency. This results in the annual loss of 1.

View Article and Find Full Text PDF

Role of Ciliary Neurotrophic Factor in Angiotensin II-Induced Hypertension.

Hypertension

January 2025

Department of Nephrology, Medical Faculty, University Hospital Düsseldorf, Heinrich Heine University Düsseldorf, Germany (S.A.P., I.Q., D. Arifaj, M.K., D. Argov, L.C.R., J.S.).

Background: Ciliary neurotrophic factor (CNTF), mainly known for its neuroprotective properties, belongs to the IL-6 (interleukin-6) cytokine family. In contrast to IL-6, the effects of CNTF on the vasculature have not been explored. Here, we examined the role of CNTF in AngII (angiotensin II)-induced hypertension.

View Article and Find Full Text PDF

Blood pressure (BP) is a dynamic measure that is frequently discussed in static terms. There exist many limitations in current documentation systems whereby documented BP values may not be reflective of the dynamic variability of BP. This study used an observational, prospective, non-randomized study design to examine the variability in BP response during intravenous vasoactive medication administration in an intensive care unit setting.

View Article and Find Full Text PDF

The aim of this study was to validate the performance of the A&D UA-1100NFC hoseless devices of two cuff sizes in monitoring blood pressure (BP) in the upper arm according to the International Organization for Standardization (ISO) 81060-2:2018/amendment (Amd) 1:2020 protocol. The accuracy of the UA-1100NFC (for arm circumferences of 22.0-32.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!