Introduction: We present our initial clinical experience using moderate deep-inspiration breath hold (mDIBH) with an active breathing control (ABC) device to reduce heart dose in the treatment of patients with early-stage, left-sided breast cancer using external beam radiation therapy (EBRT) limited to the whole breast.

Methods And Materials: Between February and August 2002, 5 patients with Stages I/II left-sided breast cancer received EBRT limited to the whole breast using an ABC device. After standard virtual simulation, patients with >2% of the heart receiving >30 Gy in free breathing were selected. All patients underwent a training session with the ABC apparatus to determine their ability to comfortably maintain mDIBH at 75% of the maximum inspiration capacity. Three patients received 45 Gy to the whole breast in 25 fractions, and 2 patients received 50.4 Gy in 28 fractions. For each of the medial and lateral tangential beams, radiation was delivered during 2 or 3 breath hold durations that ranged from 18 to 26 s. "Step-and-shoot" intensity modulation was employed to achieve uniform dose distribution. Open beam segments were purposely delivered over 2 breath hold sessions and captured on electronic portal images to allow intra- and interfraction setup error analysis. All electronic portal images of the tangential beams were analyzed off-line using an in-house treatment verification tool to assess the anteroposterior, craniocaudal, and rotational uncertainties. Corrections were applied if necessary.

Results: A comparison of treatment plans performed on breath-hold and free-breathing CTs showed that ABC treatments achieved a mean absolute reduction of 3.6% in heart volume receiving 30 Gy (heart V(30)) and 1.5% in the heart normal tissue complication probability. A total of 134 ABC treatment sessions were performed in the 5 patients. The average number of breath holds required per beam direction was 2.5 (4-6 per treatment) with a median duration of 22 s per breath hold (range: 10-26 s). Patients tolerated mDIBH well. The median treatment time was 18.2 min (range: 13-32 min), which was progressively shortened with increasing experience. A total of 509 portal images were analyzed. Combining measurements for all patients, the interfraction setup errors (1 SD) in the lateral and craniocaudal directions and in rotation were 2.4 mm, 3.2 mm, and 1 degrees, respectively, for the medial beam and 2.3 mm, 3.1 mm, and 1 degrees, respectively, for the lateral beam. For all patients, the intrafraction setup errors were about 1 mm and always less than 2 mm (1 SD).

Conclusion: Reduction in heart V(30) can be achieved in patients with left-sided breast cancer using mDIBH assisted with an ABC device. With increasing experience, ABC treatments were streamlined and could be performed within a 15-min treatment slot. Our results suggest that mDIBH using an ABC device may provide one of the most promising methods of improving the efficacy of EBRT in patients with left-sided breast cancer, particularly when wide tangential beams are employed. Breast cancer; Breath hold; Radiation therapy; Intensity modulated radiation therapy

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0360-3016(03)00010-5DOI Listing

Publication Analysis

Top Keywords

breath hold
24
breast cancer
24
left-sided breast
20
radiation therapy
16
abc device
16
patients
13
patients left-sided
12
tangential beams
12
portal images
12
initial clinical
8

Similar Publications

Background And Purpose: In lung stereotactic body radiation therapy (SBRT) using a breath-holding technique, displacement of tumor during breath-holding is rarely considered. This study used four-dimensional (4D) dose calculation with cine computed tomography (CT) to evaluate the impact of unexpected tumor position displacement during breath-holding on the target dose of lung volumetric modulated arc therapy (VMAT)-SBRT.

Materials And Methods: This study included 20 cases for which tumor position displacement during end-exhalation breath-holding (range: 0.

View Article and Find Full Text PDF

Sex-based variations in breath-holding: oxygen storage and diving response among non-divers.

Front Physiol

January 2025

Department of Health Sciences, Environmental Physiology Group, Mid Sweden University, Östersund, Sweden.

Breath-hold diving performances are typically better in men than in women. However, it is still being determined if there are differences in the physiological responses to breath-holding between the sexes. We conducted a study comparing the maximum breath-hold duration, heart rate (HR) reduction, peripheral oxygen saturation (SpO), and spleen volume and contraction in 37 men and 44 women, all of whom had no prior breath-holding experience.

View Article and Find Full Text PDF

Background: Management of respiratory motion during radiation therapy is essential for accurate dose delivery and minimizing the risk to organs. In diagnostic imaging, respiratory monitoring is required for confirmation of breath-hold and four-dimensional computed tomography (CT) reconstruction. Although respiratory monitoring systems are widely used in radiation therapy, they are not often used for diagnostic imaging, where they could improve image quality.

View Article and Find Full Text PDF

Background: Ultrasound lung surface motion measurement is valuable for the evaluation of a variety of diseases. Speckle tracking or Doppler-based techniques are limited by the loss of visualization as a tracked point moves under ribs or is dependent.

Methods: We developed a synthetic lateral phase-based algorithm for tracking lung motion to overcome these limitations.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!