Objective: Validity of a method to improve the accuracy of carotid artery duplex scanning was tested in comparison with arteriography.

Study Design: In 489 patients who had not previously undergone arteriography, 978 carotid arteries were examined with duplex ultrasound scanning. In method A, a linear array 7.5 MHz transducer with pulsed-wave 4.7 MHz Doppler scanning was used. For the diagnosis and grading of carotid stenosis, peak systolic and end-diastolic velocity of the Doppler waves were recorded. Method B consisted of complete ultrasound imaging and color-flow mapping with a convex array 3.5 MHz transducer with pulsed-wave 2.8 MHz Doppler scanning in all patients who had previously undergone method A. Further velocity measurements were performed at the sites of stenosis. The results of methods A and B were compared with data from neurologic assessment and arteriographic studies.

Results: Method B showed significantly higher diagnostic agreement with arteriography than did method A (K 95% confidence interval [CI], 0.87-0.93 vs 0.79-0.85; P <.05), and the number of mistakes in grading stenosis was significantly lower (primarily because of decreased overestimation) in patients with internal carotid kinking (>60 degrees of angulation) (P <.05), distal stenosis (>20 mm from bifurcation) (P <.01), or wide acoustic shadowing (>1 cm) (P <.01) and in those without these conditions (P <.05). Compared with arteriography, diagnostic accuracy with the new method proved higher for carotid stenoses 50% or greater, 60% or greater, 70% or greater, and 80% or greater; no statistically significant difference was found for carotid stenosis 96% or greater or for carotid occlusion. Compared with data from neurologic assessment and arteriography, method B proved more accurate than method A in designating patients for carotid endarterectomy (P =.014).

Conclusions: The new method significantly improved diagnostic reliability of duplex ultrasound scanning, especially in carotid arteries with kinking, distal stenosis, or wide acoustic shadowing (32.2% of all arteries studied). In clinical practice, we suggest additional use of a lower frequency transducer in cases in which these three conditions are found or suspected at first scanning.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0741-5214(02)75138-0DOI Listing

Publication Analysis

Top Keywords

array mhz
16
carotid artery
8
duplex ultrasound
8
ultrasound scanning
8
linear array
8
convex array
8
patients undergone
8
mhz transducer
8
transducer pulsed-wave
8
pulsed-wave mhz
8

Similar Publications

Purpose: High-frequency ultrasound (HFUS) of muscle and nerve has the potential to be a reliable, responsive, and informative biomarker of disease progression for individuals with amyotrophic lateral sclerosis (ALS). High-frequency ultrasound is not able to visualize median nerve fascicles to the same extent as ultra-high-frequency ultrasound (UHFUS). Evaluating the number and size of fascicles within a nerve may facilitate a better understanding of nerve diseases.

View Article and Find Full Text PDF

Electron spin resonance (ESR) is a powerful tool for characterizing and manipulating spin systems, but commercial ESR spectrometers can be inflexible and designed to work in narrow frequency bands. This work presents a spectrometer built from off-the-shelf parts that, when coupled with easy-to-design resonators, enables ESR over a broad frequency range, including at frequencies outside the standard bands. It can operate at either a single frequency or at two frequencies simultaneously.

View Article and Find Full Text PDF

Si APD-Based High Speed Infrared Radiation Thermometry for Analysing the Temperature Instability of a Combustion Chamber.

Sensors (Basel)

December 2024

Sensor Systems Group, School of Electrical & Electronic Engineering, The University of Sheffield, Portobello Centre, Pitt Street, Sheffield S1 4ET, UK.

This study introduces a novel approach to analysing the combustion process using a high-speed, non-contact, optical fibre-coupled Si avalanche photodiode (APD)-based infrared radiation thermometer (IRT). The Si APD-IRT, combined with an optimised field-programmable gate array (FPGA)-based digital design, achieves a response time of 1 µs, faster than commercially available instruments. Our instrument captures the entire ignition and reignition cycle of a Jet A kerosene droplet with high temporal precision within a combustion chamber, a feat impossible with traditional thermocouples.

View Article and Find Full Text PDF

Nuclear magnetic resonance relaxation of the proton spins of liquid molecules and their evolution during processes such as drying, fluid flow, and phase change of a sample can be monitored in a nondestructive way. A unilateral H NMR sensor made with a permanent magnet array, inspired by the NMR MOUSE, with an RF coil tuned to 11.71 MHz was developed.

View Article and Find Full Text PDF

Diagnostic Accuracy of Dynamic High-Resolution Ultrasonography in Assessing Anterior Disc Displacement in Temporomandibular Joint Disorders: A Prospective Observational Study.

Healthcare (Basel)

November 2024

State Key Laboratory of Oral Diseases, National Center for Stomatology, National Clinical Research Center for Oral Diseases, Department of Oral Medical Imaging, West China Hospital of Stomatology, Sichuan University, Chengdu 610041, China.

The objective of this study was to assess the diagnostic efficacy of dynamic high-resolution ultrasonography (HRUS) in detecting anterior disc displacement with reduction (ADDWR) and anterior disc displacement without reduction (ADDWoR) in the temporomandibular joint (TMJ). A total of 144 TMJs was categorized into three groups according to the magnetic resonance imaging (MRI) findings, which served as the reference standard: the normal disc position (NDP) group, the ADDWR group, and the ADDWoR group. Static images of the TMJ in full opening and maximum intercuspal positions, along with dynamic sequences during jaw opening, were obtained utilizing a 14 MHz L-shaped linear array transducer.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!