Background: ¹³C breath test analysis requires accurate ¹³CO₂measurements.
Aim: To perform a multicentre study to evaluate the repeatability and reproducibility of breath ¹³CO₂analysis.
Methods: Two series of 25 paired randomly coded tubes (each consisting of 23 ¹³CO₂-enriched breath samples and two samples of standard reference pure CO₂with certified δ ¹³C(PDB)) were sent to participating centres for ¹³CO₂measurement. Each series of tubes was analysed 10 days apart. The repeatability and reproducibility of ¹³C measurements was assessed by Mandel's k and h statistics.
Results: Twenty-two centres participated in the study: 18 showed good inter- and intra-laboratory variability, whilst four showed abnormally high inter- or intra-laboratory variability. Breath test results were also significantly affected by the accuracy of the ¹³C analytical procedures.
Conclusions: A low accuracy of ¹³C measurements may significantly affect the results of breath tests, leading to inappropriate clinical decisions. Standardization of ¹³C analysis is required to guarantee optimal ¹³C measurements and accurate ¹³C breath test results.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1046/j.1365-2036.2003.01584.x | DOI Listing |
J Thromb Haemost
January 2025
French Reference Center for Hemophilia, Hospices Civils de Lyon, Hemostasis&Thrombosis, UR4609 Universite Claude Bernard Lyon 1, Lyon, France. Electronic address:
Background: Efanesoctocog is a B-domain-deleted, Fc-fusion factor (F)VIII linked to the D'D3 domain of von Willebrand factor and 2 XTEN polypeptides, designed for an ultra-extended half-life for prophylaxis in hemophilia A, but also aiding in managing acute bleeding or surgery in patients on long-term emicizumab. However, no current laboratory method accurately measures FVIII levels in the presence of emicizumab.
Objectives: To test whether the bovine chromogenic FVIII assay, specifically calibrated for efanesoctocog, could provide an accurate assessment of efanesoctocog activity.
Electrophoresis
January 2025
Forensic Sciences Laboratory, Section of Legal Medicine, Department of Medicine and Surgery, Santa Maria Hospital, University of Perugia, Terni, Italy.
The increasing interest in DNA methylation (DNAm) analysis within the forensic scientific community prompted a collaborative project by Ge.F.I.
View Article and Find Full Text PDFEcotoxicol Environ Saf
December 2024
Aquatic Contaminants Research Division, Environment and Climate Change Canada, Montréal, Québec, Canada. Electronic address:
The Hydra vulgaris bioassay is recognized as sensitive invertebrate test species for toxicity assessment of real-life environmental mixtures for enforcement and monitoring investigations. The purpose of this study was to characterize the intra-laboratory variability, study the influence of environmental variables (temperature, luminosity, inter-individual and day of analysis) on ZnSO toxicity, a reference model toxicant for hydra. The sublethal (effect concentration for 50 % of hydra-EC50) and lethal (lethal concentration for 50 % of hydra-LC50) were determined based on characteristic morphological changes for this species.
View Article and Find Full Text PDFJ Clin Virol
December 2024
Department of Microbiology and Infection Prevention, University of Groningen. University Medical Center Groningen, Groningen, The Netherlands.
bioRxiv
October 2024
Wadsworth Center, New York State Department of Health, Albany, NY.
Dried blood spots (DBS) and oral fluids (OF) are easily attainable biospecimen types that have enabled population scale antibody monitoring for SARS-CoV-2 exposure and vaccination. However, the degree to which the two different biospecimen types can be used interchangeably remains unclear. To begin to address this question, we generated contrived DBS (cDBS) and OF (cOF) from serum panels from SARS-CoV-2 infected, vaccinated, and uninfected individuals.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!