A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Study on the accuracy of automated hematology analyzers in Shanghai. | LitMetric

Study on the accuracy of automated hematology analyzers in Shanghai.

Southeast Asian J Trop Med Public Health

Department of Clinical Hematology, Shanghai Centre for Clinical Laboratory, People's Republic of China.

Published: September 2003

Unlabelled: The objectives of this study were to research on the accuracy of automated hematology analyzers of various types from different manufacturers and to observe the deviation among these instruments. Fresh anticoagulated blood from healthy donors on 115 hematology analyzers in 114 different hospitals were determined.

Results: The maximum coefficients of variation (CVs. %) among instruments of three main manufacturers (Sysmex, Beckman Coulter and Abbott) of red blood cell count (RBC), hemoglobin (Hgb), hematocrit (Hct), white blood cell count (WBC) and platelet count (Plt) were 3.2%, 3.8%, 3.6%, 9.3% and 10.8%, respectively. The maximum deviations among these parameters of different instruments were 0.74%, 1.65%, 5.45%, 7.06% and 18.55%, respectively. By improving laboratory quality management, the results of hematology analyzer determination may be more reliable than manual methods. The difference among various manufacturers was very small about RBC, Hgb, Hct, WBC and Plt the results from all kinds of instruments will tend to be comparable.

Download full-text PDF

Source

Publication Analysis

Top Keywords

hematology analyzers
12
study accuracy
8
accuracy automated
8
automated hematology
8
blood cell
8
cell count
8
hematology
4
analyzers shanghai
4
shanghai unlabelled
4
unlabelled objectives
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!