Background: Low wall shear stress (WSS) increases neointimal hyperplasia (NH) in vein grafts and stents. We studied the causal relationship between WSS and NH formation in stents by locally increasing WSS with a flow divider (Anti-Restenotic Diffuser, Endoart SA) placed in the center of the stent.
Methods And Results: In 9 rabbits fed a high-cholesterol diet for 2 months to induce endothelial dysfunction, 18 stents were implanted in the right and left external iliac arteries (1 stent per vessel). Lumen diameters were measured by quantitative angiography before and after implantation and at 4-week follow-up, at which time, macrophage accumulation and interruption of the internal elastic lamina was determined. Cross sections of stent segments within the ARED (S+ARED), outside the ARED (S[minus]ARED), and in corresponding segments of the contralateral control stent (SCTRL) were analyzed. Changes in WSS induced by the ARED placement were derived by computational fluid dynamics. Computational fluid dynamics analysis demonstrated that WSS increased from 0.38 to 0.82 N/m2 in the S+ARED immediately after ARED placement. This augmentation of shear stress was accompanied by (1) lower mean late luminal loss by quantitative angiography ([minus]0.23+/-0.22 versus [minus]0.58+/-0.30 mm, P=0.02), (2) reduction in NH (1.48+/-0.58, 2.46+/-1.25, and 2.36+/-1.13 mm2, P<0.01, respectively, for S+ARED, S[minus]ARED, and SCTRL), and (3) a reduced inflammation score and a reduced injury score. Increments in shear stress did not change the relationship between injury score and NH or between inflammation score and NH.
Conclusions: The newly developed ARED flow divider significantly increases WSS, and this local increment in WSS is accompanied by a local reduction in NH and a local reduction in inflammation and injury. The present study is therefore the first to provide direct evidence for an important modulating role of shear stress in in-stent neointimal hyperplasia.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1161/01.CIR.0000066914.95878.6D | DOI Listing |
Comput Methods Biomech Biomed Engin
January 2025
Shenzhen Key Laboratory of Cardiovascular Disease, Fuwai Hospital Chinese Academy of Medical Sciences, Shenzhen, Guangdong Province, China.
Impeller radial gap is one of important parts within a blood pump, which may affect the hemodynamics and hemocompatibility. In this study, computational fluid dynamics method was performed to evaluate the impact of radial gap sizes. The volume of scalar shear stress decreased with radial gap sizes increasing.
View Article and Find Full Text PDFJ Clin Med
December 2024
Clinic for Masticatory Disorders and Dental Biomaterials, Center for Dental Medicine, University of Zurich, 8006 Zurich, Switzerland.
: Sinus lifting, a procedure to augment bone in the maxilla, may cause complications such as sinusitis due to impaired drainage. This study aimed to assess how sinus lifting impacts airflow in the sinus cavity, which is essential for patients undergoing dental implants. Using computational fluid dynamics (CFD), this research analyzed airflow changes after sinus floor elevation, offering insights into the aerodynamic consequences of the procedure.
View Article and Find Full Text PDFJ Clin Med
December 2024
National Laboratory Astana, Nazarbayev University, Astana 010000, Kazakhstan.
Despite the high progress that has been made in the field of cardiology, the left ventricular assist device (LVAD) can still cause complications (thrombosis/bleeding) in heart failure (HF) patients after implantation. Complications develop due to the incorrect dose of antithrombotic therapy, due to the influence of the non-physiological shear stress of the device, and also due to inherited genetic polymorphisms. Therefore, the aim of our study is to identify the influence of the genetic polymorphisms on complication development in HF patients with implanted LVADs with prescribed antiplatelet therapy.
View Article and Find Full Text PDFSensors (Basel)
December 2024
Biofluids Laboratory, Perm National Research Polytechnic University, 614990 Perm, Russia.
Simulating the cardiac valves is one of the most complex tasks in cardiovascular modeling. As fluid-structure interaction simulations are highly computationally demanding, machine-learning techniques can be considered a good alternative. Nevertheless, it is necessary to design many aortic valve geometries to generate a training set.
View Article and Find Full Text PDFMaterials (Basel)
December 2024
Suzhou XDM 3D Printing Technology Co., Ltd., Suzhou 215000, China.
The stress distribution within the struts of lattice metamaterials is non-uniform under compressive loads, with stress concentrations typically occurring at the node regions. Inspired by bamboo, this study proposes a type of body-centered cubic (BCC) lattice metamaterial with tapered prism struts (BCCT). The compressive behavior, deformation modes, mechanical properties, and failure mechanisms of BCCT lattice metamaterials are systematically analyzed using finite element methods and validated through compression tests.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!