Simplified sample workup obviating protein precipitation and eluent evaporation commonly employed in earlier reports using chloroformate-mediated derivatization of aminothiols prior to mass spectrometric (MS) detection is presented. The reduction of disulfides in plasma is accomplished with dithiothreitol within minutes. A simultaneous derivatization with ethyl chloroformate (ECF) and extraction of derivatives into organic phase takes place within seconds. Along with S-amino acids, also aromatic amino acids can be determined during a 5-min run. Gas chromatography with flame ionization detection (GC-FID) proved to be sensitive enough to reach plasma homocysteine levels. A prerequisite for a reliable quantitation was fulfilled under the given conditions. Intra-assay precision was <5%, recoveries from spiked plasma complete (101.2%), detection and quantitation limits for homocysteine came to <1 and 3 micro mol/l. Our results were in full agreement with those obtained by liquid chromatography (r=0.999 for homocysteine and 0.987 for cysteine), and were close to two homocysteine immunoassays (r=0.991 and 0.939, respectively).

Download full-text PDF

Source
http://dx.doi.org/10.1016/s1570-0232(03)00104-1DOI Listing

Publication Analysis

Top Keywords

aromatic amino
8
amino acids
8
simple plasma
4
plasma work-up
4
work-up fast
4
fast chromatographic
4
chromatographic analysis
4
analysis homocysteine
4
homocysteine cysteine
4
cysteine methionine
4