Background: The 940-nm diode laser has been shown to be an effective treatment for leg veins.

Objective: We sought to evaluate the effectiveness of the 940-nm diode laser on leg veins, depending on the size and morphologic aspect of the treated vessels.

Methods: A total of 60 patients (mean age: 44.4 years, Fitzpatrick skin types I-IV) underwent up to 3 treatment sessions at 4-week intervals using the 940-nm diode laser. Treatment parameters were: vessels < 0.4 mm in diameter, 0.5-mm spot, pulse duration of 10 milliseconds, fluence 306 J/cm(2); 0.4 to 0.8 mm in diameter, 1-mm spot, pulse duration of 30 milliseconds, fluence 306 J/cm(2); and 0.8 to 1.4 mm in diameter, 1.5-mm spot, pulse duration of 70 milliseconds, fluence 317 J/cm(2). Repetition rate was 2.5 Hz. Success rate was evaluated through double-blind observation.

Results: Only 13.33% of patients with telangiectases less than 0.4 mm in diameter had a percentage of vessel clearance superior to 75%. However, 88.24% of patients with vessels between 0.8 and 1.44 mm in diameter obtained more than 75% vessel clearance.

Conclusion: The treatment of leg veins by the 940-nm diode laser strongly depends on the size of the target vessel. Better results were obtained with 0.8- to 1.4-mm leg venulectases.

Download full-text PDF

Source
http://dx.doi.org/10.1067/mjd.2003.191DOI Listing

Publication Analysis

Top Keywords

diode laser
20
940-nm diode
16
treatment leg
12
spot pulse
12
pulse duration
12
duration milliseconds
12
milliseconds fluence
12
laser effective
8
effective treatment
8
leg veins
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!