Objectives: Salivary analysis is a noninvasive alternative that may be more acceptable to patients, especially children. As such, it has the potential for incorporation into comprehensive, dynamic investigations of metabolic dysfunctions - a significant advancement over a single time point serum analysis. In this study, we wanted to determine if the serum cortisol assay on our routine immunoassay analyzer could reliably measure salivary cortisol concentrations. Because of potential fluctuations in salivary concentrations, we included a biologic variation study as a main facet of our preliminary method evaluation.

Design And Methods: Twenty-eight healthy individuals (12 males, 16 females) volunteered to provide 5 nonconsecutive first morning saliva samples over a two-week period. Samples were stored frozen at home until the completion of the study. Following thawing and centrifugation, cortisol was measured in batch mode for each set of participant samples on the ROCHE Elecsys. Biologic variation was determined following removal of outliers. A method comparison was performed with the DPC Coat-A-Count Cortisol assay following the recommended modifications for salivary analysis, and with the Salimetrics HS-Cortisol two-site monoclonal assay optimized for salivary cortisol.

Results: Mean salivary cortisol concentration was 20.4 nmol/L. Analytical variation (CV(A) = 3.8%), within-subject variation (CV(I) = 6.3%), between-subject variation (CV(G) = 20.5%), index of individuality (II = 0.36) and reference change value (RCV = 20.4%) were determined. A negative 40% proportional bias was observed on the Elecsys compared to the two methods that have already been optimized for salivary analysis.

Conclusions: This study demonstrated that salivary cortisol can be reliably measured on a routine automated immunoassay analyzer such as the ROCHE Elecsys. This particular assay needs to be optimized at the low end of the standard curve for routine use with salivary samples. Based on the relatively small intra-individual variation and low index of individuality, reference change values are preferable to the use of population reference intervals for this assay.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0009-9120(02)00471-xDOI Listing

Publication Analysis

Top Keywords

salivary cortisol
16
roche elecsys
12
salivary
10
salivary analysis
8
cortisol assay
8
immunoassay analyzer
8
biologic variation
8
assay optimized
8
optimized salivary
8
reference change
8

Similar Publications

Background: The period before diagnostic and therapeutic procedures is associated with increased anxiety levels in patients due to a lack of sufficient information. This study aimed to determine the effect of education on physiological and psychological anxiety levels in patients before extracorporeal shock wave lithotripsy (ESWL).

Materials And Methods: This randomized clinical trial was conducted at Baharloo Hospital in Tehran on 122 patients, and the samples were selected in two stages.

View Article and Find Full Text PDF

Blood Flow Restricted Resistance Exercise in Well-Trained Men: Salivary Biomarker Responses and Oxygen Saturation Kinetics.

J Strength Cond Res

December 2024

Jayhawk Athletic Performance Laboratory, Wu Tsai Human Performance Alliance, University of Kansas, Lawrence, Kansas.

Eserhaut, DA, DeLeo, JM, and Fry, AC. Blood flow restricted resistance exercise in well-trained men: Salivary biomarker responses and oxygen saturation kinetics. J Strength Cond Res 38(12): e716-e726, 2024-Resistance exercise with continuous lower-limb blood flow restriction (BFR) may provide supplementary benefit to highly resistance-trained men.

View Article and Find Full Text PDF

Introduction: Colonoscopy is currently the most commonly used and effective method for early detection, diagnosis and treatment of tumours of the colon and rectum. However, similar to other invasive procedures, it is associated with adverse reactions such as pain and abdominal distension. Electroacupuncture (EA) has been proposed as a potential treatment for relieving this discomfort; however, there is limited evidence supporting its efficacy.

View Article and Find Full Text PDF

Introduction: Cortisol is an essential stress hormone and failure of its production, known as adrenal insufficiency (AI), is associated with significant mortality due to adrenal crisis. The Short Synacthen Test (SST) is the current diagnostic test of choice for AI, but it is both invasive and resource intensive. Globally, there is an unmet need for a non-invasive, cost-effective test.

View Article and Find Full Text PDF

.

Compr Psychoneuroendocrinol

November 2024

School of Social Sciences, University of Westminster, London, UK.

Cortisol awakening response (CAR) research relies upon self-collected saliva sampling during the post-awakening period. It is unknown how the CAR protocol is perceived and how they may affect typical routines relevant to CAR methodology. CAR assessment is sensitive to state variables, suggesting that CAR measurement may be affected by research participation.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!