We have developed an intravascular thermography basket catheter to measure the temperature of the vessel wall to locate foci of inflammation. Our 3 Fr thermography basket catheter is a thermocouple-based catheter made of a nitinol expandable and externally controllable basket system loaded with nine small and flexible built-in thermosensors. It is equipped with real-time data acquisition software with a thermal resolution of 0.0001 degrees C and a sampling rate of 20 readings per second. In 10 inbred cholesterol-fed dogs with femoral (but not carotid) atherosclerosis, we found foci of warmth on the surface of atherosclerotic but not disease-free regions (P < 0.05). Marked temperature heterogeneity was also observed in the aortas of atherosclerotic Watanabe rabbits but not in normal rabbits. The catheters showed satisfactory accuracy, reproducibility, and safety. If confirmed in further studies, it has the potential to be utilized in detection of vulnerable plaques.

Download full-text PDF

Source
http://dx.doi.org/10.1002/ccd.10486DOI Listing

Publication Analysis

Top Keywords

thermography basket
12
basket catheter
12
detection vulnerable
8
vulnerable plaques
8
catheter
4
catheter vivo
4
vivo measurement
4
measurement temperature
4
temperature atherosclerotic
4
atherosclerotic plaques
4

Similar Publications

Article Synopsis
  • Researchers investigated how different laser pulse modes affect temperature changes during laser lithotripsy using thermography.
  • An artificial kidney model was tested with four modes: short pulse, long pulse, virtual basket, and Moses mode, focusing on temperature areas above 43°C over 60 seconds.
  • The study found that the long pulse and Moses modes created larger high-temperature areas compared to the short pulse and virtual basket modes, which could help in reducing thermal injuries during renal surgeries.
View Article and Find Full Text PDF

This study aims to evaluate changes in irrigation fluid temperatures during laser activation by using thermography, with comparison between Moses mode (MM) and virtual basket mode (VBM). Experiments were performed using an unroofed pyelocaliceal model. The laser was fired for 60 seconds at 0.

View Article and Find Full Text PDF

We have developed an intravascular thermography basket catheter to measure the temperature of the vessel wall to locate foci of inflammation. Our 3 Fr thermography basket catheter is a thermocouple-based catheter made of a nitinol expandable and externally controllable basket system loaded with nine small and flexible built-in thermosensors. It is equipped with real-time data acquisition software with a thermal resolution of 0.

View Article and Find Full Text PDF

In 1996, we showed that inflamed atherosclerotic plaques give off more heat and that vulnerable plaques may be detected by measuring their temperature. Plaque temperature is correlated directly with inflammatory cell density and inversely with the distance of the cell clusters from the luminal surface. It is inversely related to the density of the smooth muscle cells.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!