Intra-aortic balloon pumping in acute mitral regurgitation reduces aortic impedance and regurgitant fraction.

Shock

Departments of Cardio Thoracic Surgery and Cardiology, Cardiovascular Research Institute Maastricht, Academic Hospital, Maastricht, The Netherlands.

Published: April 2003

Acute mitral regurgitation (MR) is present in 10% of patients presenting with cardiogenic shock. To stabilize these patients, intra-aortic balloon pumping (IABP) is recommended, but the mechanism of IABP support in these patients is unknown. This animal study was designed to describe the hemodynamic effect of intra-aortic balloon pumping during cardiogenic shock induced by acute MR. In eight calves, left ventricular pressure-volume loops, aortic and left atrial pressure, and aortic, carotid artery, and coronary blood flow were recorded. Acute MR (range 36%-79%) was created by placing a metal cage in the mitral valve. Hemodynamic data was obtained at control, during acute MR, and during acute MR with 1:1 IABP support. Acute MR caused a decrease in cardiac output (-32%, P = 0.018), blood pressure, and carotid artery flow, whereas left ventricular output (+127%, P = 0.018), end-diastolic volume, and left atrial pressure all significantly increased. Stroke work, ejection fraction, and coronary blood flow were not significantly changed, and no signs of ischemia were seen on the ECG. The IABP raised average cardiac output by 31% (P = 0.012) and significantly raised blood pressure and flow to the brain while decreasing systemic vascular resistance. Left ventricular function and mean coronary blood flow did not change, but diastolic coronary flow became more important as shown by the increase in diastolic fraction from 64% to 95%. (P = 0.028). Average MR dropped by 7.5% (P = 0.025). In conclusion, application of the IABP during acute MR lowers aortic impedance, resulting in less MR and more output toward the aorta without changing left ventricular function.

Download full-text PDF

Source
http://dx.doi.org/10.1097/00024382-200304000-00007DOI Listing

Publication Analysis

Top Keywords

left ventricular
16
intra-aortic balloon
12
balloon pumping
12
coronary blood
12
blood flow
12
acute
8
acute mitral
8
mitral regurgitation
8
aortic impedance
8
cardiogenic shock
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!