Purpose: This study compares the percentage decrease in power output of light-emitting diode (LED) and quartz-tungsten-halogen (QTH) curing devices at different distances between the light guide tip and a radiometer.
Materials And Methods: Three LED curing devices (Elipar FreeLight/3M ESPE, Luxomax/Akeda, e-Light/GC) were compared to two QTH curing devices (Elipar TriLight/3M ESPE, Optilux 501 conventional and Turbo light guides/Kerr-Demetron). Power density was measured with a Fieldmaster GS/Coherent unit (Sensor LM-3 HTD) at increasing distances from the light guide tip (0 to 20 mm at 1-mm increments, n=6).
Results: The mean decrease in power density available for curing at a distance of 10 mm between the light guide tip and the radiometer was 68% for the FreeLight, 83% for the e-Light, 42% for the Luxomax, 38% for the TriLight, 33% for the Optilux with the standard light guide, and 44% for the Optilux with the turbo light guide. The power density of the Luxomax was only 40% of that of the other LED curing devices at 0 mm distance.
Conclusion: The blue LED curing devices Elipar FreeLight and GC e-Light showed a significantly higher percentage decrease in power output at a distance of 10 mm from the light guide tip to the radiometer compared to the QTH curing devices. Therefore, although blue LED curing devices might have the same curing potential compared to a QTH device when placed in direct contact to a resin composite, blue LED curing devices may not provide a sufficient cure when placed at a clinically relevant distance of 10 mm to the resin composite surface.
Download full-text PDF |
Source |
---|
Polymers (Basel)
December 2024
Science and Technology on Reliability Physics and Application of Electronic Component Laboratory, China Electronic Product Reliability and Environmental Testing Research Institute, Guangzhou 511370, China.
During the thermal aging process of epoxy resin, microcracks, interfacial delamination, and warpage are the key factors leading to semiconductor device damage. Here, epoxy-resin specimens (EP-Ss) and epoxy-resin/silicon-wafer composites (EP-SWs) were prepared to analyze the distribution of residual stress (RS) in epoxy resin and its thermal aging process changes. The uniaxial tensile approach and Raman spectroscopy (RAS) showed that the peak shift of aliphatic C-O in EP-Ss was negatively correlated with the external stress, and that the stress correlation coefficient was -2.
View Article and Find Full Text PDFNanomaterials (Basel)
December 2024
School of Mechanical and Ocean Engineering, Mokpo National University, Muan-gun 58554, Jeollanam-do, Republic of Korea.
Superhydrophobic surfaces, known for their exceptional water-repellent properties with contact angles exceeding 150°, are highly regarded for their effectiveness in applications including self-cleaning, antifouling, and ice prevention. However, the structural fragility and weak durability of conventional coating limit their long-term use. In this research, a new approach is proposed for the fabrication of long-lasting superhydrophobic surfaces using ethyl cyanoacrylate (ECA) and a primer.
View Article and Find Full Text PDFJ Contemp Dent Pract
September 2024
Department of Academic, Faculty of Dentistry, Universidad Nacional Federico Villarreal, Lima, Peru, ORCID: https://orcid.org/0000-0002-0594-5834.
Objective: To evaluate the shear strength of adhesives based on the type of solvent (ethanol and acetone), aged and light-cured using light-emitting diode (LED) units with different wavelengths. Polywave and monowave LED units were employed for this study.
Materials And Methods: Ninety bovine tooth samples were analyzed using OptiBond Universal adhesive (acetone) and single bond universal adhesive (ethanol).
Low Urin Tract Symptoms
January 2025
Department of Urology, Institute of Urology (Laboratory of Reconstructive Urology), West China Hospital, Sichuan University, Chengdu, Sichuan, People's Republic of China.
Aim: This study aimed to analyze the clinical efficacy and safety of autologous fascial pubovaginal sling (AFPVS) surgery in treating recurrent stress urinary incontinence (SUI) following the failure of mid-urethral sling procedures.
Methods: A retrospective analysis was conducted on the clinical data of SUI patients who underwent AFPVS at our hospital between 2008 June and 2024 June following the failure of mid-urethral sling procedures. The analysis included basic information, surgical parameters, and postoperative complications.
Acta Neurochir (Wien)
January 2025
Division of Neurosurgery, Beth Israel Deaconess Medical Center, Harvard Medical School, 110 Francis Street , Boston, MA, 02215, USA.
Background: Variability in long-term endovascular treatment outcomes for intracranial aneurysms has prompted questions regarding the effects of these treatments on aneurysm hemodynamics. Endovascular techniques disrupt aneurysmal blood flow and shear, but their influence on intra-aneurysmal pressure remains unclear. A better understanding of aneurysm pressure effects may aid in predicting outcomes and guiding treatment decisions.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!