Objective: To qualitatively validate an enzyme immunoassay to measure leukotriene B4 in exhaled breath condensate. Exhaled breath condensate is a new non-invasive method to monitor airway inflammation.

Subjects: Twenty-two subjects with different lung diseases attended the outpatient clinic on one occasion for exhaled breath condensate collection.

Methods: Samples were pooled together and purified by reverse-phase high-performance liquid chromatography. The fractions eluted were assayed for leukotriene B4 by enzyme immunoassay.

Results: A single peak of leukotriene B4-like immunoreactivity co-eluting with leukotriene B4 standard (retention time: 24 min) was identified by enzyme immunoassay. Reverse phase-high performance liquid chromatography peak of leukotriene B4 was clearly separated from those of 6-trans-leukotriene B4 (retention time: 14 min) and leukotriene B5 (retention time: 18 min) for which the antiserum used in the enzyme immunoassay had the highest cross-reactivity. Leukotriene B4 recovery was 64%.

Conclusions: This study provides evidence for the presence of leukotriene B4 in the exhaled breath condensate and the specificity of the enzyme immunoassay used.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s000110300003DOI Listing

Publication Analysis

Top Keywords

exhaled breath
20
breath condensate
20
enzyme immunoassay
16
retention time
12
time min
12
leukotriene
8
leukotriene exhaled
8
liquid chromatography
8
peak leukotriene
8
exhaled
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!