Blood-drawing tubes made from plastic containing K2EDTA as an anticoagulant are gaining widespread use in clinical hematology. We compared complete blood count parameters, reticulocyte counts, white blood cell differentials, and flagging rates obtained with Becton Dickinson Vacutainer K3EDTA glass tubes and Vacutainer K2EDTA Plus plastic tubes and found only slight discrepancies in the results obtained with the two tube types. Although some parameter values obtained with K3EDTA glass tubes were significantly lower than those obtained with K2EDTA plastic tubes, many of these differences could be explained by the known effects of the liquid K3EDTA anticoagulant. Flagging rates on an automated cell counter were identical for the two tube types. We conclude that the differences between results obtained with K3EDTA glass tubes versus K2EDTA plastic tubes are minimal and unlikely to be of any clinical significance.