Background And Purpose: Body contours can potentially be used for patient set-up verification in external-beam radiotherapy and might enable more accurate set-up of patients prior to irradiation. The aim of this study is to test the feasibility of patient set-up verification using a body contour scanner.

Material And Methods: Body contour scans of 33 lung cancer and 21 head-and-neck cancer patients were acquired on a simulator. We assume that this dataset is representative for the patient set-up on an accelerator. Shortly before acquisition of the body contour scan, a pair of orthogonal simulator images was taken as a reference. Both the body contour scan and the simulator images were matched in 3D to the planning computed tomography scan. Movement of skin with respect to bone was quantified based on an analysis of variance method.

Results: Set-up errors determined with body-contours agreed reasonably well with those determined with simulator images. For the lung cancer patients, the average set-up errors (mm)+/-1 standard deviation (SD) for the left-right, cranio-caudal and anterior-posterior directions were 1.2+/-2.9, -0.8+/-5.0 and -2.3+/-3.1 using body contours, compared to -0.8+/-3.2, -1.0+/-4.1 and -1.2+/-2.4 using simulator images. For the head-and-neck cancer patients, the set-up errors were 0.5+/-1.8, 0.5+/-2.7 and -2.2+/-1.8 using body contours compared to -0.4+/-1.2, 0.1+/-2.1, -0.1+/-1.8 using simulator images. The SD of the set-up errors obtained from analysis of the body contours were not significantly different from those obtained from analysis of the simulator images. Movement of the skin with respect to bone (1 SD) was estimated at 2.3 mm for lung cancer patients and 1.7 mm for head-and-neck cancer patients.

Conclusion: Measurement of patient set-up using a body-contouring device is possible. The accuracy, however, is limited by the movement of the skin with respect to the bone. In situations where the error in the patient set-up is relatively large, it is possible to reduce these errors using a computer-aided set-up technique based on contour information.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0167-8140(02)00369-9DOI Listing

Publication Analysis

Top Keywords

patient set-up
24
simulator images
24
body contours
20
body contour
16
cancer patients
16
set-up errors
16
set-up
12
lung cancer
12
head-and-neck cancer
12
movement skin
12

Similar Publications

In chronic lymphocytic leukemia, the reliability of next-generation sequencing (NGS) to detect variants ≤10% allelic frequency (low-VAF) is debated. We tested the ability to detect 23 such variants in 41 different laboratories using their NGS method of choice. The sensitivity was 85.

View Article and Find Full Text PDF

Background: Ureteral stents, such as double-J stents, have become indispensable in urologic procedures but are associated with complications like hematuria and pain. While the advancement of artificial intelligence (AI) technology has led to its increasing application in the health sector, AI has not been used to provide information on potential complications and to facilitate subsequent measures in the event of such complications.

Objective: This study aimed to assess the effectiveness of an AI-based prediction tool in providing patients with information about potential complications from ureteroscopy and ureteric stent placement and indicating the need for early additional therapy.

View Article and Find Full Text PDF

Low frequency noise (LFN) of unknown origin can be disturbing, especially at night, and affect sleep and relaxation. If reduction of the LFN is impossible, adding sound can mask the troubling noise or detract attention from it to reduce distress. To assess the effectiveness of this, a survey was set up consisting of three questionnaires: the first when a person requested a masking sound and follow-ups after 1 and 3 months.

View Article and Find Full Text PDF

Background: National stroke clinical quality registries/audits support improvements in stroke care. In a 2016 systematic review, 28 registries were identified. Since 2016 there have been important advances in stroke care, including the development of thrombectomy services.

View Article and Find Full Text PDF

Purpose: To determine the dosimetric effects of set-up errors on boost coverage, and compares skin toxicity of sequential and simultaneous boost techniques for left-sided breast cancer.

Materials And Methods: This retrospective study included 23 early-stage breast cancer cases. Single isocenter HFWBI-SIB(s-SIB), single isocenter HFWBI-SB(s-SB) and dual isocenter HFWBI-SB(d-SB) were planing.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!