Reliability of pachymetric measurements using orbscan after excimer refractive surgery.

Ophthalmology

Fondation Ophtalmologique Adolphe de Rothschild, Hôpital Bichat, 25 à 29 Rue Manin, 75019 Paris, France.

Published: March 2003

Purpose: To assess the accuracy of pachymetric measurements using Orbscan (Bausch & Lomb, Rochester, NY) after laser in situ keratomileusis (LASIK) and photorefractive keratectomy (PRK).

Design: Prospective instrument validation study.

Participants: Seventy-nine nonoperated normal eyes, 84 eyes after LASIK, and 50 eyes after PRK.

Intervention: Laser in situ keratomileusis or PRK.

Methods: Central corneal thickness was measured using ultrasound and Orbscan II. The acoustic factor (AF) was adjusted, based on the results obtained in the normal eye group, to minimize the difference between ultrasound and Orbscan pachymetric values.

Main Outcome Measures: Central corneal thickness as measured by Orbscan and ultrasound pachymeter.

Results: Using the adjusted AF, which was 0.946, the mean difference between Orbscan and ultrasonic pachymetric measurements was 0 +/- 17, 16 +/- 28, and 68 +/- 39 microm in the normal, LASIK, and PRK groups, respectively. The difference between all groups was statistically significant (P < 0.0001).

Conclusions: Orbscan pachymetric values may be underestimated and less accurate after LASIK and PRK.

Download full-text PDF

Source
http://dx.doi.org/10.1016/S0161-6420(02)01298-8DOI Listing

Publication Analysis

Top Keywords

pachymetric measurements
12
measurements orbscan
8
laser situ
8
situ keratomileusis
8
central corneal
8
corneal thickness
8
thickness measured
8
ultrasound orbscan
8
orbscan pachymetric
8
+/- +/-
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!