Purpose: Development and experimental evaluation of an intravascular monitoring system for telemetric measurement of blood pressure and heart rate.

Materials And Methods: The monitoring system consists of an implantable silicone capsule (diameter 2.3 mm), containing a dedicated microchip with pressure sensors and signal-processing circuits as well as an antenna for wireless data and energy transfer using 6.78 MHz transponder technology. Three self-expanding legs at one end of the capsule served as a mechanism to lock the capsule at an arterial branch. A flow model, driven by a ventricular assist system, was used for testing and optimizing the implantation equipment, for checking the anchoring mechanism and for ensuring transmission of the measured pressure to the readout unit. In-vivo experiments were performed in 8 minipigs (weight 25 to 30 kg), with three capsules placed in each minipig via the femoral artery using a dedicated 8-F sheath/pusher system. Follow-up was by CT angiography for up to 6 months after implantation.

Results: Flow model tests revealed a maximum deviation of pressure and heart rate measurements of 5% from the reference measurements. Signal transmission was reliable over a distance of 3 to 4 cm. Fluoroscopically guided in-vivo implantation of the capsules was simple and straightforward. In arteries with a diameter of 5 to 6 mm, the capsules were permanently fixed with one or two legs interlocked in side branches and without occlusion within 6 months. Three capsules developed a small non-occlusive appositional thrombus attached to the downstream (leg) part of the capsule.

Conclusion: Our in-vitro and in-vivo experiments demonstrate the feasibility of wireless transmission from a capsule with a sufficient resolution of the sensor output signals to determine blood pressure and pulse rate. As long as the vessel diameter is wide enough, arterial fixation of the capsule does not induce thrombotic occlusion of the parent artery. With respect to future clinical applications, further refinements of the transmission technology are needed to extend the transmission distance between capsule and reader antenna. The technology of intelligent implants has further implications, such as monitoring of other physiological parameters, and the design of a control loop, which may be used for therapeutic feedback.

Download full-text PDF

Source
http://dx.doi.org/10.1055/s-2003-37232DOI Listing

Publication Analysis

Top Keywords

blood pressure
12
monitoring system
8
pressure heart
8
flow model
8
in-vivo experiments
8
three capsules
8
pressure
6
capsule
6
transmission
5
vascular capsule
4

Similar Publications

Arterial NALCN Knock-Down Ameliorates Mineralocorticoid-Induced Hypertension and Arterial Over-Contractility.

Circ Res

January 2025

Department of Physiology, Institute of Functional Genomics and Research Institute of Medical Science, Konkuk University School of Medicine, Chungju, Republic of Korea (H.L., S.P., J.R.A., M.S.S., H.J.N., B.K., Y.M.B.).

View Article and Find Full Text PDF

Background: The arterial stiffening is attributed to the intrinsic structural stiffening and/or load-dependent stiffening by increased blood pressure (BP). The respective lifetime alterations and major determinants of the two components with normal aging are not clear.

Methods: A total of 3053 healthy adults (1922 women) aged 18-79 years were enrolled.

View Article and Find Full Text PDF

Introduction: Hypertension is the leading preventable cause of cardiovascular morbidity and mortality globally, with a disproportionate impact on low-income and middle-income countries like Sri Lanka. Effective blood pressure (BP) control improves outcomes in patients with hypertension. This study aimed to assess the prevalence of uncontrolled hypertension, and its correlates among Sri Lankan patients with hypertension in clinic settings.

View Article and Find Full Text PDF

A 37-year-old man presented with symptoms of polyuria and weight loss over the past year. Initial laboratory examination showed elevated blood glucose level (468 mg/dL [25.9 mmol/L]; normal reference range [RR], 75-109 mg/dL [4.

View Article and Find Full Text PDF

Decoding morbidity patterns in ASHA workers of Vadodara, Gujarat - A cross-sectional study.

J Family Med Prim Care

December 2024

Department of Community Medicine, Smt. B.K. Shah Medical Institute and Research Centre, Sumandeep Vidyapeeth Deemed to be University, Piparia, Vadodara, Gujarat, India.

Introduction: Accredited Social Health Activists (ASHA) play a pivotal role as village-level functionaries under the National Health Mission, driving community transformation, facilitating health planning, and ensuring access to healthcare services.

Objectives: To assess the physical health status of ASHA workers and to investigate the prevalence of morbidities among ASHA workers.

Material And Methods: This is a cross-sectional study, conducted in Vadodara's east zone, and involves 123 ASHA workers.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!