Use of internal coils for independent and direct MR imaging-guided endovascular device tracking.

J Vasc Interv Radiol

Department of Radiology, Northwestern University Medical School, 676 North St. Clair, Suite 800, Chicago, Illinois 60611, USA.

Published: February 2003

Purpose: To test the hypotheses that a single internal guide wire coil (i) permits independent and direct depiction of guide wires and catheters and (ii) improves catheter-tracking accuracy and depiction compared to external receiver coils.

Materials And Methods: Standard 5-6-F angiographic catheters were filled with dilute 4% gadolinium chelate. A single 0.030-inch-diameter internal guide wire coil was placed inside the catheter. True fast imaging with steady-state precession was used to directly visualize the guide wire. Inversion recovery-prepared fast low-angle shot technique was used to track catheters over a thick slice. In phantom experiments, we compared catheter signal-to-noise ratios (SNRs) with the internal coil and a phased-array surface coil with use of the Wilcoxon signed-rank test. Tip-tracking accuracy was assessed with use of linear regression. In pigs (n = 7), catheters and guide wires were independently tracked in real time.

Results: In phantoms, catheter SNR with the internal coil (12.0) was significantly greater than that with the surface coil (4.0; P =.001). Tip-tracking accuracy was also improved with use of the internal coil (R(2) = 0.94 vs 0.50). In swine vasculature, catheters and guide wires could be directly and independently tracked at 1.7-2.0 frames per second. Catheters were clearly visualized with use of the internal coil, with a typical catheter background contrast-to-noise ratio of 6.6. Catheters were not visible with use of the external coil because of the small catheter size compared to the slice thickness.

Conclusion: Internal guide wire coils permit independent and direct depiction of guide wires and catheters in vivo for MR imaging-guided endovascular interventions. They also improve catheter tracking accuracy and depiction compared to external coils.

Download full-text PDF

Source
http://dx.doi.org/10.1097/01.rvi.0000058328.82956.15DOI Listing

Publication Analysis

Top Keywords

guide wire
16
guide wires
16
internal coil
16
independent direct
12
internal guide
12
coil
9
internal
8
imaging-guided endovascular
8
guide
8
wire coil
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!