Assessment of external fixator reusability using load- and cycle-dependent tests.

Clin Orthop Relat Res

Orthopaedic Biomechanics Laboratory, School of Medicine, Johns Hopkins University, 720 Rutland Avenue, Baltimore, MD 21205-2196, USA.

Published: January 2003

No standard method has been established for investigating repeated use of an external fixator. The purpose of the current study was to establish a fatigue testing method for assessing fixator frame reuse. A unilateral DynaFix trade mark external fixator system was tested using high-load and low-cycle (900-150 N at 5 Hz) and low-load and high-cycle (450-100 N at 10 Hz) tests (assumed one use of 500,000 and 1 million cycles, respectively). These loading conditions were selected to simulate single clinical use and to satisfy Food and Drug Administration requirements. In the high-load low-cycle test, substantial failure of the serrated joint occurred before completion of the first simulated use. In the low-load high-cycle test, all fixators completed three simulated clinical uses without failure, although (1/4) of the serrated joint components had hairline cracks. The high-load low-cycle test identified the fixator components which should be examined and replaced if reuse of the fixator is to be considered. Wear and deformation of the set screw on the rotary joint and telescoping mechanisms were observed in the low-load high-cycle test but not in the high-load low-cycle test. Therefore, if the unilateral DynaFix trade mark fixators are being considered for reusability, the number of reuses should be limited as the whole structure of the device will experience fatigue damage as the loading cycle increases.

Download full-text PDF

Source
http://dx.doi.org/10.1097/01.blo.0000037443.23683.7cDOI Listing

Publication Analysis

Top Keywords

high-load low-cycle
16
external fixator
12
low-load high-cycle
12
low-cycle test
12
unilateral dynafix
8
dynafix trade
8
trade mark
8
serrated joint
8
high-cycle test
8
fixator
6

Similar Publications

Design considerations for piezoelectrically powered electrical stimulation: The balance between power generation and fatigue resistance.

J Mech Behav Biomed Mater

February 2022

University of Kansas, Bioengineering Graduate Program, Lawrence, KS, USA; Evoke Medical, LLC, Lawrence, KS, USA; University of Kansas, Department of Mechanical Engineering, Lawrence, KS, USA. Electronic address:

Quality and timing of bone healing from orthopedic surgeries, especially lumbar spinal fusion procedures, is problematic for many patients. To address this issue, clinicians often use electrical stimulation to improve surgery success rates and decrease healing time in patients with increased risk of pseudarthrosis, including smokers and diabetics. Current invasive electrical stimulation devices require an implantable battery and a second surgery for removal.

View Article and Find Full Text PDF

Influence of CAD/CAM milling, sintering and surface treatments on the fatigue behavior of lithium disilicate glass ceramic.

J Mech Behav Biomed Mater

January 2021

School of Mechanical Engineering, The University of Adelaide, Adelaide, SA, 5005, Australia. Electronic address:

This paper reports on the process-fatigue relation of lithium disilicate glass ceramic (LDGC) using low-cycle, high-load Hertzian indentations with a rigid indenter to simulate teeth grinding/clenching of LDGC restorations with different surface asperities obtained in CAD/CAM milling, sintering, polishing and glazing. The maximum contact stresses were evaluated as functions of the number of load cycles and surface treatments using the Hertzian model. Indentation-induced surface damage was viewed using scanning electron microscopy (SEM) to understand the relationships among microstructures, surface asperities, crack morphology and propagation.

View Article and Find Full Text PDF

Assessment of external fixator reusability using load- and cycle-dependent tests.

Clin Orthop Relat Res

January 2003

Orthopaedic Biomechanics Laboratory, School of Medicine, Johns Hopkins University, 720 Rutland Avenue, Baltimore, MD 21205-2196, USA.

No standard method has been established for investigating repeated use of an external fixator. The purpose of the current study was to establish a fatigue testing method for assessing fixator frame reuse. A unilateral DynaFix trade mark external fixator system was tested using high-load and low-cycle (900-150 N at 5 Hz) and low-load and high-cycle (450-100 N at 10 Hz) tests (assumed one use of 500,000 and 1 million cycles, respectively).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!