Inhaled nitric oxide fraction is influenced by both the site and the mode of administration.

J Clin Monit Comput

Département d'Anesthésie-Réanimation & SMUR, Assistance Publique-Hôpitaux de Paris, Hôpital Lariboisière, 2 rue Ambroise Paré, 75475 Paris Cedex 10, France.

Published: December 1999

Objective: Inhaled nitric oxide (NO) can be delivered continuously or sequentially (= during inspiration) at different locations of the ventilation circuit. We have tested the influence of locations, modes of NO administration and the ratio of the inspiratory time over the respiratory cycle time (I/I + E ratio) on the accuracy of NO fractions, delivered by 2 devices: Opti-NO and Flowmeter.

Methods: We used a simplified lung model consisting of a ventilation circuit with a Y piece, a tracheal tube, a 150 ml dead-space volume and a 5 liter balloon. Three fractions (3, 6, 9 ppm) were administered continuously or sequentially, in controlled volume, in 4 different sites on the inspiratory branch above the Y piece: i) just after the water trap, ii) just before the Y piece; below the Y piece: iii) just after the Y piece, iv) into the endotracheal tube. In addition, different I/I + E ratios (25, 33, 50, 80%) were studied. The delivered NO fractions were measured in the balloon by chemiluminescence (CLD 700, Ecophysics). A linear regression analysis was used to test the relationship between administered and measured NO fractions for the 3 fractions (3, 6 and 9 ppm) in sequential and continuous modes. Intercept values were compared to zero and slopes to the identity line.

Results: When NO was administered in the continuous mode upstream the Y piece, NO fractions measured in the balloon corresponded to the administered fractions. In contrast, below the Y piece, the measured NO fractions were significantly lower than the administered NO fractions. In the sequential mode, above and below the Y piece, the delivered NO fractions were within the manufacturer's range.

Conclusions: For the continuous NO delivery, locations above the Y piece are mandatory. However, locations below the Y piece imposes a sequential system, which can also be used for the sites located above the Y piece.

Download full-text PDF

Source
http://dx.doi.org/10.1023/a:1009971712989DOI Listing

Publication Analysis

Top Keywords

piece
11
fractions
10
inhaled nitric
8
nitric oxide
8
continuously sequentially
8
ventilation circuit
8
fractions ppm
8
delivered fractions
8
fractions measured
8
measured balloon
8

Similar Publications

Background: Central venous access devices (CVAD) are widely used in patient care, providing an essential, reliable pathway for patients to receive chemotherapy, long-term infusions, and nutritional support. However, a system of exercise management has not been developed in patients with CVAD.

Purpose: To evaluate and summarize the evidence for management exercise in patients with CVAD and provide guidance for clinical practice.

View Article and Find Full Text PDF

Background: This paper aimed to investigate the correlation between estradiol and testosterone in patients with breast cancer.

Methods: The research papers on the correlation between estradiol and testosterone on the risk of breast cancer were searched and collected. The time limit is that each database was established until December 2023.

View Article and Find Full Text PDF

The promise of community-driven preprints in ecology and evolution.

Proc Biol Sci

January 2025

Evolution and Ecology Research Centre, School of Biological, Earth and Environmental Sciences, University of New South Wales, Sydney, New South Wales 2052, Australia.

Publishing preprints is quickly becoming commonplace in ecology and evolutionary biology. Preprints can facilitate the rapid sharing of scientific knowledge establishing precedence and enabling feedback from the research community before peer review. Yet, significant barriers to preprint use exist, including language barriers, a lack of understanding about the benefits of preprints and a lack of diversity in the types of research outputs accepted (e.

View Article and Find Full Text PDF

We report a high peak power mid-infrared (MIR) source via efficient optical parametric generation (OPG) in a piece of 50-mm-long MgO: PPLN crystal pumped by using a near-infrared (NIR) narrow-band picosecond laser source. The highest peak power of the idler pulse can reach ∼109.86 kW with a duration of ∼7.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!