The evaluation of tonometry and self-tonometry with Ocuton tonometers.

Med Sci Monit

1st Department of Ophthalmology, Semmelweis University, Budapest, Hungary.

Published: January 2003

Background: The purpose of the present study was to evaluate the clinical applicability of the Ocuton-A and Ocuton-S applanation tonometers, and to compare their use and accuracy parameters to those of Goldmann applanation tonometry (GAT).

Material/methods: In the first study, intraocular pressure (IOP) was measured with an Ocuton-A tonometer, followed by another measurement using GAT in 15 subjects (30 eyes). Ocuton-tonometry was performed on three occasions separated by three-minute intervals. In Study 2 the impact of increased patient familiarity on measurement accuracy using the self-tonometer was investigated in 5 subjects (10 eyes) by comparing two consecutive series of five Ocuton-S measurements. In Study 3 nine trained volunteers measured their own IOP with an Ocuton-S self-tonometer after GAT and Ocuton-A measurements by an experienced investigator.

Results: Study 1 showed that the measured IOP value was significantly higher using the Ocuton-A tonometer, compared to values obtained using GAT. In Study 2 we found that increased practice in self-tonometry did not alter measurement accuracy. In Study 3 the difference between IOP readings obtained with the Ocuton-S and Ocuton-A instruments was not significant.

Conclusions: The portable Ocuton-A tonometer may become a useful instrument for IOP measurement by professional workers outside the clinic. Qualified patients are able to use the Ocuton-S self-tonometer reliably even after limited training, and its measurement accuracy is acceptable for IOP monitoring. Both the Ocuton A and S devices consistently overestimate the IOP by approximately 2 mm Hg compared to the corresponding Goldmann readings.

Download full-text PDF

Source

Publication Analysis

Top Keywords

ocuton-a tonometer
12
measurement accuracy
12
subjects eyes
8
measured iop
8
ocuton-s self-tonometer
8
study
7
iop
7
ocuton-a
6
ocuton-s
5
measurement
5

Similar Publications

Background: Golmann applanation tonometry represents a well-established procedure for measuring intraocular pressure (IOP). This implies the necessity of an accurate measurement of IOP with the reference tonometer. One example is the contour tonometer Pascal with a measuring probe, adapted to the cornea geometry, for measuring the IOP and the ocular pulse amplitude.

View Article and Find Full Text PDF

Background: The purpose of the present study was to evaluate the clinical applicability of the Ocuton-A and Ocuton-S applanation tonometers, and to compare their use and accuracy parameters to those of Goldmann applanation tonometry (GAT).

Material/methods: In the first study, intraocular pressure (IOP) was measured with an Ocuton-A tonometer, followed by another measurement using GAT in 15 subjects (30 eyes). Ocuton-tonometry was performed on three occasions separated by three-minute intervals.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!