Background: The purpose of the present study was to evaluate the clinical applicability of the Ocuton-A and Ocuton-S applanation tonometers, and to compare their use and accuracy parameters to those of Goldmann applanation tonometry (GAT).
Material/methods: In the first study, intraocular pressure (IOP) was measured with an Ocuton-A tonometer, followed by another measurement using GAT in 15 subjects (30 eyes). Ocuton-tonometry was performed on three occasions separated by three-minute intervals. In Study 2 the impact of increased patient familiarity on measurement accuracy using the self-tonometer was investigated in 5 subjects (10 eyes) by comparing two consecutive series of five Ocuton-S measurements. In Study 3 nine trained volunteers measured their own IOP with an Ocuton-S self-tonometer after GAT and Ocuton-A measurements by an experienced investigator.
Results: Study 1 showed that the measured IOP value was significantly higher using the Ocuton-A tonometer, compared to values obtained using GAT. In Study 2 we found that increased practice in self-tonometry did not alter measurement accuracy. In Study 3 the difference between IOP readings obtained with the Ocuton-S and Ocuton-A instruments was not significant.
Conclusions: The portable Ocuton-A tonometer may become a useful instrument for IOP measurement by professional workers outside the clinic. Qualified patients are able to use the Ocuton-S self-tonometer reliably even after limited training, and its measurement accuracy is acceptable for IOP monitoring. Both the Ocuton A and S devices consistently overestimate the IOP by approximately 2 mm Hg compared to the corresponding Goldmann readings.
Download full-text PDF |
Source |
---|
Klin Monbl Augenheilkd
September 2012
Background: Golmann applanation tonometry represents a well-established procedure for measuring intraocular pressure (IOP). This implies the necessity of an accurate measurement of IOP with the reference tonometer. One example is the contour tonometer Pascal with a measuring probe, adapted to the cornea geometry, for measuring the IOP and the ocular pulse amplitude.
View Article and Find Full Text PDFMed Sci Monit
January 2003
1st Department of Ophthalmology, Semmelweis University, Budapest, Hungary.
Background: The purpose of the present study was to evaluate the clinical applicability of the Ocuton-A and Ocuton-S applanation tonometers, and to compare their use and accuracy parameters to those of Goldmann applanation tonometry (GAT).
Material/methods: In the first study, intraocular pressure (IOP) was measured with an Ocuton-A tonometer, followed by another measurement using GAT in 15 subjects (30 eyes). Ocuton-tonometry was performed on three occasions separated by three-minute intervals.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!