Background: The distal-balloon protection system is being evaluated for its efficacy in preventing embolic neurological events during carotid stenting (CAS). We sought to determine the effect of this system on the frequency of the Doppler-detected microembolic signals (MES) during CAS.
Methods: Using transcranial Doppler, we compared the frequency of MES during CAS in 2 groups; 39 patients without distal protection and 37 with the distal-balloon protection system (GuardWire, Percusurge, Sunnyville, CA). There was no significant difference in the clinical or angiographic characteristics between the 2 groups. Three phases with increased MES counts were identified during the unprotected CAS; stent deployment, predilation, and postdilation (mean+/-SD: 75+/-57, 32+/-36, and 27+/-25, respectively).
Results: The distal-balloon protection significantly reduced the frequency of MES during CAS (MES-counts: 164+/-108 in the control vs 68+/-83 in the protection group, p=0.002) particularly during these 3 phases. MES in the protection group were detected predominantly during sheath placement, guidewire manipulation and during the distal-balloon deflation.
Conclusions: Three phases with increased MES counts were identified during the unprotected CAS, e.g. stent deployment, predilation and postdilation. The distal-balloon protection system significantly reduced the frequency of MES during CAS, particularly during these 3 phases.
Download full-text PDF |
Source |
---|
AJNR Am J Neuroradiol
May 2024
Department of Neurosurgery (K.Y., M. Hayakawa, K.A., Y.H.), Fujita Health University, Toyoake, Aichi, Japan.
Background And Purpose: Tumor embolization through the meningohypophyseal trunk and inferolateral trunk is known to be effective in skull base tumors; however, microcatheter cannulation into these arteries is difficult, and the number of cases that can be safely embolized is limited. In this study, we present a novel embolization procedure for the meningohypophyseal trunk and inferolateral trunk using the distal balloon protection technique and detail its clinical efficacy and complication risks. We developed this procedure to allow safe embolization in patients who cannot be adequately cannulated with microcatheters into these arteries.
View Article and Find Full Text PDFCardiovasc Revasc Med
August 2023
HM CIEC, HM Hospitals, Madrid, Spain; CEU San Pablo University, Spain; Cardiac Imaging Unit, Spain.
No-reflow phenomenon is frequent in patients with ST-segment elevation myocardial infarction (STEMI) and has proven to be a strong predictor of mortality. Local fibrinolytic infusion with distal coronary occlusion (previously described as "marinade technique") can be useful in patients with acute myocardial infarction and intraluminal thrombus refractory to aspiration enabling the local effect of the drug, directly applied inside the thrombus, while protecting the microvasculature with prolonged inflation of a distal balloon. We present the early experience of four patients with inferior acute myocardial infarction and high thrombus burden successfully treated with marinade technique in one center.
View Article and Find Full Text PDFVascular
April 2024
Division of Vascular Surgery and Endovascular Therapy, University of Colorado, Anschutz Medical Center, Aurora, CO, USA.
Objectives: Carotid artery stenting (CAS) is an alternative treatment option for patients at high risk for carotid endarterectomy (CEA) but has been correlated with increased risk for distal embolization and periprocedural stroke despite the use of adjunctive embolic protection devices (EPD). This study compared four types of EPDs and their intra and periprocedural related complications.
Methods: A systematic review of the literature was conducted in PubMed/Medline to identify studies that investigated the outcomes of CAS with adjuvant use of EPDs, including Proximal Balloon (PB), Distal Filter (DF), and Distal Balloon (DB) strategy.
World Neurosurg
December 2022
Department of Neurosurgery, Shinshu University School of Medicine, Shinshu University Hospital, Matsumoto, Japan; Neuroendovascular Therapy Center, Shinshu University Hospital, Matsumoto, Japan.
Background: Embolic protection devices are commonly used to prevent cerebral embolism during carotid artery stenting (CAS). However, the PercuSurge Guardwire device, a distal balloon protection device, was discontinued in April 2021. We present a case of successful CAS using a new distal balloon protection system.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!