Transluminal stent graft repair with Wallgraft endoprosthesis in a porcine arteriovenous graft pseudoaneurysm model.

J Vasc Surg

Division of Vascular Surgery & Endovascular Therapy, the Michael E. DeBakey Department of Surgery, Baylor College of Medicine, Houston VAMC (112), 2002 Holcomb Blvd, Houston, TX 77030, USA.

Published: January 2003

Purpose: Pseudoaneurysm is a known complication of arteriovenous grafts in chronic hemodialysis and can result in graft disruption or thrombosis if left untreated. This study evaluated the safety and efficacy of endovascular repair with Wallgraft endoprosthesis (Boston Scientific, Inc, Watertown, Mass) in a porcine arteriovenous graft (AVG) pseudoaneurysm model.

Materials And Methods: Bilateral groin AVG pseudoaneurysms (n = 18) were created with an oversized Dacron interposition graft within a polytetrafluoroethylene femoral AVG in nine domestic swine and allowed to mature 28 +/- 4 days (standard deviation). Transluminal placement of Wallgraft was performed to exclude the pseudoaneurysm from the AVG circulation. Hemodialysis was performed (400 mL/min x 1 hour, with intravenous heparin 30 units/kg) every 4 days for a total of 6 weeks via 15-gauge needles in the treated AVG pseudoaneurysm site. Arteriography and duplex ultrasound scan were performed to determine AVG patency and pseudoaneurysm flow. Histologic evaluation was performed to determine Wallgraft morphology. In vitro pulsatile flow chamber was used to determine maximal flow volume without peri-Wallgraft endoleak.

Results: All AVG pseudoaneurysms were successfully excluded with the Wallgrafts. Twelve AVG (67%) remained patent at the completion of the study. No Wallgraft migration occurred from hemodialysis. Transient peri-Wallgraft endoleak (<2 hours after hemodialysis) was present in 13 of 18 (72%) and four of 12 (33%) AVG pseudoaneurysms by weeks 1 and 6, respectively. With maintenance of an intraluminal pressure of 80, 100, 120, 140, and 160 mm Hg in the pulsatile flow chamber, the maximal flow rates without peri-Wallgraft endoleak were 625 +/- 120, 650 +/- 145, 620 +/- 95, 425 +/- 110, and 262 +/- 86 mL/min. Scanning electron microscopy showed a neointimal layer covered with thrombus on the Wallgraft surface.

Conclusion: Endoluminal placement of Wallgraft endoprosthesis provides adequate structural support for continuous hemodialysis after AVG pseudoaneurysm exclusion. Transient blood flow in the pseudoaneurysm cavity may occur immediately after the hemodialysis, which may represent the effect of heparin used during hemodialysis. This study suggests Wallgraft is a safe and effective treatment for AVG pseudoaneurysm and permits continuous hemodialysis.

Download full-text PDF

Source
http://dx.doi.org/10.1067/mva.2002.87DOI Listing

Publication Analysis

Top Keywords

repair wallgraft
8
wallgraft endoprosthesis
8
porcine arteriovenous
8
arteriovenous graft
8
avg
8
avg pseudoaneurysm
8
avg pseudoaneurysms
8
performed determine
8
pseudoaneurysm
6
graft
5

Similar Publications

Objective Endovascular stent graft repair for exclusion of pseudoaneurysm is currently being explored to replace open surgical repair as current management. The objective of the study was to evaluate the clinical safety and efficacy of endovascular stent graft intervention in pseudoaneurysms complicating arteriovenous reconstructions in patients on chronic hemodialysis. Methods A retrospective analysis of all pseudoaneurysms treated with stent grafts at our institution over a period of five years was performed.

View Article and Find Full Text PDF

Background: Injury of the ureter is a potential complication of any difficult abdominopelvic surgical procedure, with an incidence ranging between 0.5% and 10% in most series. The treatment depends on the severity and the place of the lesion.

View Article and Find Full Text PDF

Endovascular repair of aorto-iliac artery injuries after lumbar-spine surgery.

Eur J Vasc Endovasc Surg

August 2011

Department of Vascular and Thoracic Surgery, Arnaud de Villeneuve Hospital, Montpellier, France.

Objective: This study aims to describe the endovascular management of abdominal-aortic- or common-iliac-artery injuries after lumbar-spine surgery.

Methods: Patients treated for abdominal-aortic- or common-iliac-artery injuries after lumbar-spine surgery during a 13-year period were identified from an endovascular database, providing prospective information on techniques and outcome. The corresponding patient records and radiographic reports were analysed retrospectively.

View Article and Find Full Text PDF

Purpose: To evaluate the feasibility of endovascular exclusion of popliteal artery aneurysm (PAA) using stent grafts.

Methods: The clinical data of all patients who underwent endovascular exclusion of PAA at three French vascular departments between December 1999 and December 2007 were retrospectively analyzed. Outcome measures included graft patency and endoleak.

View Article and Find Full Text PDF

Purpose: To determine the safety and efficacy of percutaneous treatment for popliteal artery aneurysms (PAA) using self-expanding stent-grafts.

Methods: From October 2000 through September 2007, 29 patients (27 men; mean age 68+/-6 years, range 54-88) underwent endovascular exclusion of 33 PAAs. All had symptoms of claudication, and all were atherosclerotic in origin.

View Article and Find Full Text PDF