Under controlled imaging conditions, functional, co-registered ventilation and perfusion images were obtained for quantitative MRI assessment of pulmonary function. Experiments were conducted on 25-35 kg Yorkshire pigs (N = 5) under ketamine/isoflurane anesthesia. Ventilation and perfusion MR images were obtained using laser polarized gas and gadolinium (Gd), respectively. The mutual information-based technique was used to quantify the initial misregistration between the acquired ventilation and perfusion images, to cope with different intensity characteristics. Means and standard deviations (SDs) of misregistrations for the five subjects, in terms of absolute translation and absolute rotations around each of the three coordinate axes, were computed to evaluate the hypothesis. For qualitative illustrations, plots of mutual information values under absolute translations and rotations around each of the three coordinate axes are presented. This preliminary study shows the feasibility of acquiring functional image data that are spatially co-registered. This method has potential applications in the quantitative assessment of regional alveolar ventilation and pulmonary perfusion ratio, in both normal and diseased lung.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mrm.10359DOI Listing

Publication Analysis

Top Keywords

ventilation perfusion
16
perfusion images
12
acquired ventilation
8
rotations three
8
three coordinate
8
coordinate axes
8
ventilation
5
perfusion
5
co-registration acquired
4
perfusion images--validation
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!