An episode of device related proarrhythmia observed with a dual chamber ICD is reported. Ventricular tachycardia was induced due to an inappropriate interaction of programmed device settings with automatic algorithms, including noncompetitive atrial pacing (NCAP), an algorithm used to prevent atrial tachyarrhythmias.