Background: The use of intensity modulated radiation fields needs an extended quality assurance concept. This consists of a linac related part and a case related part. Case related means the verification of an individual treatment plan, optimized on a CT data set of an individual patient and prepared for the treatment of this patient. This part of the quality assurance work is usually time consuming, delivers only partially quantitative results and is uncomfortable without additional help. It will be shown in this paper how the software VERIDOS will improve the optimization of the case related part of the quality assurance work.

Material And Methods: The main function of the software is the quantitative comparison of the calculated dose distribution from the treatment planning software with the measured dose distribution of an irradiated phantom. Several additional functions will be explained. Two self-developed phantoms made of RW3 (solid water) and GAFCHROMIC films or Kodak EDR2 films for the measurement of the dose distributions were used. VERIDOS was tested with the treatment planning systems Helay-TMS and Brainscan.

Results: VERIDOS is a suitable tool for the import of calculated dose matrices from the treatment planning systems Helax-TMS and Brainscan and of measured dose matrices exported from the dosimetry software Mephysto (PTW). The import from other treatment planning systems and scanning software applications for film dosimetry is generally possible. In such case the import function has to be adapted to the special header of the import matrix. All other functions of this software tool like normalization (automatically, manually), working with corrections (ground substraction, factors), overlay/comparison of dose distributions, difference matrix, cutting function (profiles) and export functions work reliable.

Conclusions: VERIDOS improves the optimization of the case related part of the quality assurance work for intensity modulated radiation therapy (IMRT). The diverse functions of the software offer the radiation physicist a wide base to verify the IMRT plan independent from the mode of its delivery (compensator technology or MLC technology).

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00066-002-0986-8DOI Listing

Publication Analysis

Top Keywords

quality assurance
20
treatment planning
16
intensity modulated
12
planning systems
12
modulated radiation
8
assurance work
8
optimization case
8
case quality
8
calculated dose
8
dose distribution
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!