Background: Dry powder inhalers (DPI) have in recent years become a common mode for administration of inhaled corticosteroids for preventive therapy of asthma. Inhaled steroids delivered by DPI achieve increased lung deposition compared with pressurized metered-dose inhalers (pMDI), which is associated with increased therapeutic effect. This may be associated with increased systemic absorption.

Objective: The purpose of this study was to evaluate the prevalence of adrenal suppression in children using low-dose budesonide given by DPI, as compared with pMDI attached to a large-volume spacer device (pMDI + spacer).

Methods: In an open-labeled crossover study, 15 asthmatic children aged 5 to 15 years received 200 microg of inhaled budesonide twice daily by DPI (Turbuhaler, Astra, Draco AB, Lund, Sweden) and by pMDI + spacer, 1 month each, in a randomized order. Twenty-four-hour urine collections were performed at baseline and at the end of each of the 2 months of the study period, and urinary cortisol and creatinine were measured.

Results: Baseline urinary cortisol:creatinine was 0.038 +/- 0.012 microg/mg, similar in both groups. After 1 month of DPI therapy, urinary cortisol:creatinine was reduced by 27 +/- 16% to 0.028 +/- 0.012 microg/mg (P = 0.018). Urinary cortisol:creatinine after 1 month of pMDI + spacer therapy was similar to baseline 0.037 +/- 0.019 microg/mg (P = 0.78).

Conclusions: Treatment of asthmatic children with budesonide 400 microg daily given via a DPI for 1 month was associated with hypothalamic-pituitary-adrenal axis suppression. This effect was not observed with the same dose of budesonide administered via pMDI + spacer. This indicates that systemic absorption might be reduced with pMDI + spacer therapy.

Download full-text PDF

Source
http://dx.doi.org/10.1016/S1081-1206(10)62103-5DOI Listing

Publication Analysis

Top Keywords

pmdi spacer
16
asthmatic children
12
urinary cortisolcreatinine
12
adrenal suppression
8
inhaled budesonide
8
dry powder
8
pressurized metered-dose
8
associated increased
8
daily dpi
8
+/- 0012
8

Similar Publications

The objective of this study was to assess the use of pMDI alone and pMDI with different spacers in asthmatic patients and to identify any associations between errors in handling the device for the first time and the sessions needed to reach the correct handling method, considering patient demographics and clinical characteristics. A total of 150 Asthmatic patients were crossed over to handle pMDI alone and with add-on inhalable devices (Aerochamber plus, Tips Haler, Able, Dispozable and Aer-8) randomly, without receiving verbal or demonstrative instruction (baseline assessment). The assessment of the inhaler technique was performed using checklists that had been set beforehand.

View Article and Find Full Text PDF

Several techniques had been developed to generate aerosolized medications during noninvasive ventilation (NIV) using variable inhalation methods. This study hypothesized that large spacers were more efficient significantly than small spacers and adapters during NIV. The main objective of this study was to compare the performance of newly developed spacers with standard T-piece in NIV chronic obstructive pulmonary disease (COPD) subjects.

View Article and Find Full Text PDF

Background: There is a scarcity of high-quality research on the efficient delivery of inhaled corticosteroids using valved holding chambers (VHCs) in children.

Methods: The delivered dose (DD) of fluticasone from a metered dose inhaler (pMDI) was tested using four VHCs: AeroChamber plus Flow-Vu (AC), Babyhaler (BH), EasyChamber (EC), and Optichamber Diamond (OD). The in vitro setup included an anatomical child throat model, Next Generation Impactor, and a breathing simulator to generate tidal breathing of a four and a 6-year-old child, and adult type single inhalation.

View Article and Find Full Text PDF

Comparative performance study of paperboard disposable spacers versus commercial valved holding chambers for aerosol delivery.

Int J Pharm

December 2024

Mines Saint-Etienne, Univ Jean Monnet, INSERM, U 1059 Sainbiose, Centre CIS, F - 42023 Saint-Etienne France. Electronic address:

Purpose: The aim of this study is to evaluate and compare the performance, for the administration of fluticasone propionate with a pressurized metered-dose inhaler (pMDI), of two low-tech paperboard spacers versus two commercially available valved holding chambers (VHC).

Methods: According to the Canadian standard CAN/CSA-Z264.1-02, total emitted dose (TED) and aerodynamic size distribution were measured for the pMDI in combination with 4 different spacers: a homemade paper cup spacer, the DispozABLE® paperboard spacer, the AeroChamber Plus® plastic VHC, and the Vortex® aluminium VHC.

View Article and Find Full Text PDF

Small Airways Disease Affects Aerosol Deposition in Children with Severe Asthma: A Functional Respiratory Imaging Study.

J Aerosol Med Pulm Drug Deliv

December 2024

Department of Pediatrics, Division of Respiratory Medicine and Allergy, Erasmus MC-Sophia Children's Hospital, University Medical Center Rotterdam, Rotterdam, The Netherlands.

Small airways disease (SAD) in severe asthma (SA) is associated with high disease burden. Effective treatment of SAD could improve disease control. Reduced end-expiratory flows (forced expiratory flow [FEF] and FEF) are considered sensitive indicators of SAD.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!