Objectives: To examine the validity of the "Gaehwiler" actigraph (Gaehwiler Electronics, model Z80-32k V(1)) for the assessment of sleep-wake (S/W) rhythm and sleep structure in infants during the first 6 months of life using an algorithm developed in our laboratory to differentiate sleep and wake states.
Methods: A continuous 72 h actigraphic recording was performed in 10 healthy infants at 1, 3 and 6 months of age. The actigraphic data were matched to direct observation of the infants' behavioural states. Using discriminant function analysis a scoring algorithm for automatic identification of S/W states from raw activity data was developed. The chi-square periodogram analysis was performed to estimate periodic components of S/W rhythm.
Results: The overall agreement rates between the actigraphic and observer scoring for S/W were between 87 and 95% for the infants after the third month of life, while for the 1-month-old infants they never exceeded 72%. The actigraphic discrimination between active and quiet sleep was the best in 3-month-old infants. The circadian influence on S/W rhythm was already present by the end of the first month of life.
Conclusions: Using the "Gaehwiler" actigraph in our study, valid discrimination between sleep and wake states was obtained in infants during 3 and 6 months. The actigraph, however, did not provide valid active vs. quiet sleep state measures. The circadian rhythm of S/W was observed as early as during the first month of life.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/s1388-2457(02)00287-0 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!