The determination of O(2) consumption by using arteriovenous O(2) content differences is dependent on accurate oxyhemoglobin saturation measurements. Because swine are a common experimental species, we describe the validation of CO-oximeter for porcine-specific oxyhemoglobin saturation. After developing a nonlinear mathematical model of the porcine oxyhemoglobin saturation curve, we made 366 porcine oxyhemoglobin saturation determinations with a calibrated blood-gas analyzer and a porcine-specific CO-oximeter. There was a high degree of correlation with minimal variability (r(2) = 0.99, SE of the estimate = 5.2%) between the mathematical model and the porcine-specific CO-oximeter measurements. Bland-Altman comparison showed that the CO-oximeter measurements were biased slightly lower (-0.4 vol%), and the limits of agreement (+/-2 SD) were 0.7 and -1.5 vol%. This is in contrast to a 10-20 vol% error if human-specific methods were used. The results show excellent agreement between the nonlinear model and CO-oximeter for porcine-specific oxyhemoglobin saturation measurements. In contrast, comparison of the porcine-specific oxyhemoglobin saturations with saturations obtained by using human methods highlights the necessity of species-specific measurement methodology.

Download full-text PDF

Source
http://dx.doi.org/10.1152/japplphysiol.00710.2002DOI Listing

Publication Analysis

Top Keywords

oxyhemoglobin saturation
20
saturation measurements
12
porcine-specific oxyhemoglobin
12
co-oximeter porcine-specific
8
mathematical model
8
porcine oxyhemoglobin
8
porcine-specific co-oximeter
8
co-oximeter measurements
8
porcine-specific
6
saturation
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!