Purpose: To model the pressure relationship between an intraocular lens (IOL) and the capsular bag.
Setting: Department of Physics, Kings' College, London University, London, United Kingdom.
Methods: A mathematical model was made of the forces between the capsule and IOL showing that the pressure related to the local radius of curvature of the IOL at any given point. The local radius of curvature for round-edged and square-edged IOLs was measured from electron micrographs of the IOL profiles, and the corresponding pressure profiles were calculated and compared.
Results: The pressure between an IOL and the capsular bag was proportional to the quotient of the tension in the capsule divided by the local radius of curvature of the IOL, with a constant of proportionality that depended on the coefficient of friction between the capsule and IOL. Measuring the local radius of curvature of the 2 IOL types suggested a pressure increase of at least 69% +/- 6% at the optic edge with the square-edged IOL.
Conclusions: The mathematical model predicted that IOLs with square-edged optic profiles exerts higher pressure on the posterior capsule than IOLs with round-edged optic profiles. The higher pressure may form a physical barrier to lens epithelial cell migration onto the posterior capsule.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/s0886-3350(02)01490-6 | DOI Listing |
J Acoust Soc Am
January 2025
Acoustic Technology, Department of Electrical & Photonics Engineering, Technical University of Denmark, Kongens Lyngby, Denmark.
Characterising acoustic fields in rooms is challenging due to the complexity of data acquisition. Sound field reconstruction methods aim at predicting the acoustic quantities at positions where no data are available, incorporating generalisable physical priors of the sound in a room. This study introduces a model that exploits the general time structure of the room impulse response, where a wave-based expansion addresses the direct sound and early reflections, localising their apparent origin, and kernel methods are applied to the late part.
View Article and Find Full Text PDFJ Orthop Case Rep
January 2025
Department of Orthopaedics, Sri Ramachandra Institute of Higher Education and Research, Chennai, Tamil Nadu, India.
Introduction: Giant cell tumour or osteoclastoma is benign, locally aggressive tumor with bone destruction and with malignant potential. It accounts for 5% of all primary bone tumor and occurs in skeletally mature individuals in the age group of 30 to 45 with peak incidence in the 3rd decade. GCT is more common in females.
View Article and Find Full Text PDFInt Ophthalmol
January 2025
Department of Ophthalmology, Staedtisches Klinikum Dessau, Brandenburg Medical School Theodor Fontane, Dessau, Germany.
Purpose: Uveal melanoma (UM) is the most common primary ocular malignancy. The size and location of the tumor are decisive for brachytherapy with the β-emitting ruthenium-106 (Ru-106) plaque. The treatment of juxtapapillary and juxtafoveolar UM may be challenging because of the proximity or involvement of the macula and optic nerve and high recurrence rates.
View Article and Find Full Text PDFPatient Saf Surg
January 2025
Department of Trauma, University Hospital Zurich, Raemistrasse 100, Zurich, 8091, Switzerland.
Introduction: Regional anesthesia increases in popularity in orthopaedic surgery. It is usually applied in elective surgeries of the extremities. The aim of this study was to assess indication of the use of general anesthesia in the surgical treatment of distal radius fractures.
View Article and Find Full Text PDFUltrasonics
January 2025
Department of Civil Engineering and Architecture, Tallinn University of Technology, Ehitajate tee 5, 19086, Tallinn, Estonia. Electronic address:
Pipe bends are recognized as critical areas susceptible to wall thinning, a phenomenon instigated by abrupt changes in the fluid flow direction and velocity. Conventional monitoring techniques for bends typically depend on localized ultrasonic measurements of thickness. While these methods are effective, they can be time-consuming compared to the use of permanently installed transducers, a strategy employed in guided wave tomography (GWT).
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!