Spiral CT evaluation of pulmonary emphysema using a low-dose technique.

Radiol Med

Radiologia Zompatori, Policlinico S. Orsola-Malpighi, Bologna, Italy.

Published: December 2002

Purpose: To evaluate the diagnostic accuracy and clinical acceptability of low-dose spiral CT for determining pulmonary volumes and emphysema extension in patients with pulmonary emphysema, in comparison with studies based on spiral CT at conventional dose.

Materials And Methods: We prospectively evaluated eighteen patients, current or former smokers, with a clinical diagnosis of chronic obstructive pulmonary disease. All the patients underwent: HRCT with three scans at predetermined levels; quantitative spiral CT, with two inspiratory scans, one conventional scan at 240 mA, and the second one a low-dose scan at 80 mA. We used the following parameters: 120 kV, rotation time 0.8", scan time less than 20" (single inspiratory breath-hold), layer thickness 7.5 mm, pitch 6 (high speed), interpolation algorithm at 180 degrees. A 3D reconstruction was performed, with segmentation of the lungs and automatic quantification of pulmonary volumes. We compared the volumes of absolute and percent emphysema and the ratings of the dose delivered to the patient (CTDIw and DLP) obtained with the two spiral CT scans with each other and with the respiratory function tests.

Results: The average total lung capacity (TLC) obtained by conventional-dose spiral CT (CTs1) was 6889.4 cc (SD +/-1813.2), and the capacity with low-dose spiral CT (CTs2) was 6929.4 cc (SD +/-1811.6). The percentage of emphysema was 39.7% (range: 2.2-63.5%; SD: +/-19.9) for the CTs1 and 41.1% (range: 2.1-66.4%; SD: +/-20). The CTDIw corresponding to CTs1 was 12.2 mGy (range: 11.9-16.4; SD: +/-1), the one corresponding to CTs2, 3.6 mGy (range: 3.6-4.9; SD: +/-0.3). The DLP corresponding to CTs1 was 391.7 mGy x cm (range: 333.3-518.9; SD: +/-46.7), the one corresponding to CTs2 was 117.8 mGy x cm (range: 100.3-156; SD: +/-14). As for the respiratory function tests, the total lung capacity (TLC) obtained by body plethysmography was 7061 cc (SD: +/-2029.7); the percent TLC was 115.9 (range: 66-165; SD: +/-27.6), the forced expiratory volume at one second (FEV1%, percentage of predicted value) was 46.7% (range: 17-123; SD: +/-27.3), residual volume (RV%) as a percentage of predicted value was 186.3 (range: 84-359; SD: +/-80.7), the Tiffeneau index (TI) was 46% (range: 25-71; SD: +/-15.7). We observed a very significant correlation between radiological and functional TLC for both CT methods. The percentage scores for emphysema obtained with the two methods correlated significantly with the functional indexes. The pixel index of CTs1 correlated with TLC% (r=0.87; p<0.0001), FEV1% (r=-0.53; p<0.02), RV% (r=0.76; p=0.004), TI (r=-0.79; p=0.0001). The pixel index of CTs2 correlated with TLC% (r=0.87; p<0.0001), FEV1% (r=-0.56; p=0.01), RV% (r=0.78; p=0.003), TI (r=-0.8; p=0.0001). The adoption of the method with low tube current entailed a highly significant reduction in the estimated dose delivered to patients (CTDIw and DLP) with r=0.9 and p < 0.0001.

Discussion And Conclusions: Quantitative low-dose spiral CT is a very good method to quantify pulmonary volumes and calculate the extension of the anatomic emphysema. The reduction of mA from 240 to 80 lowers the estimated dose by 30%, without compromising the accuracy of the results. Our study achieved a highly significant correlation between the results obtained with the two spiral CT techniques and between these results and the respiratory function tests. In clinical practice, the easiest way to reduce the dose in spiral CT of the lung is to reduce the tube current. The low-dose method allows a significant reduction in radiation exposure. Further studies are required to establish to what extent the dose can be reduced without increasing in quantum noise and thereby compromising the quality of the study.

Download full-text PDF

Source

Publication Analysis

Top Keywords

mgy range
16
range
10
pulmonary emphysema
8
low-dose spiral
8
pulmonary volumes
8
respiratory function
8
total lung
8
lung capacity
8
capacity tlc
8
corresponding cts1
8

Similar Publications

Purpose: This study evaluated beam quality and radiation dosimetry of a CT scanner equipped with a novel detector and filtration technology called PureVision Optics (PVO). PVO features miniaturized electronics, a detector cut with microblade technology, and increased filtration in order to increase x-ray detection and reduce image noise.

Methods: We assessed the performance of two similar 320-detector CT scanners: one equipped with PVO and one without.

View Article and Find Full Text PDF

Intraindividual Comparison of Image Quality Between Low-Dose and Ultra-Low-Dose Abdominal CT With Deep Learning Reconstruction and Standard-Dose Abdominal CT Using Dual-Split Scan.

Invest Radiol

January 2025

From the Department of Radiology, Ulsan University Hospital, Ulsan, Republic of Korea (T.Y.L.); Department of Radiology, University of Ulsan College of Medicine, Seoul, Republic of Korea (T.Y.L.); Department of Radiology, Seoul National University Hospital, Seoul, Republic of Korea (J.H.Y., H.K., J.M.L.); Department of Radiology, Seoul National University College of Medicine, Seoul, Republic of Korea (J.H.Y., S.H.P., J.M.L.); Department of Radiology, Inje University Busan Paik Hospital, Busan, Republic of Korea (J.Y.P.); Department of Radiology, Seoul National University Bundang Hospital, Seongnam, Republic of Korea (S.H.P.); Department of Radiology, Hanyang University College of Medicine, Seoul, Republic of Korea (C.L.); Division of Biostatistics, Medical Research Collaborating Center, Seoul National University Hospital, Seoul, Republic of Korea (Y.C.); and Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul, Republic of Korea (J.M.L.).

Objective: The aim of this study was to intraindividually compare the conspicuity of focal liver lesions (FLLs) between low- and ultra-low-dose computed tomography (CT) with deep learning reconstruction (DLR) and standard-dose CT with model-based iterative reconstruction (MBIR) from a single CT using dual-split scan in patients with suspected liver metastasis via a noninferiority design.

Materials And Methods: This prospective study enrolled participants who met the eligibility criteria at 2 tertiary hospitals in South Korea from June 2022 to January 2023. The criteria included (a) being aged between 20 and 85 years and (b) having suspected or known liver metastases.

View Article and Find Full Text PDF

Free-Breathing Respiratory Triggered High-Pitch Lung CT: Insights From Phantom and Patient Scans.

Invest Radiol

January 2025

From the Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany (A. Schwarz, A. Simon, A.M.); Siemens Healthineers AG, Forchheim, Germany (A. Schwarz, C.H., J.D., A. Simon); Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany (F.K.W., S.G., M.S.); and Institut for Radiology, Pediatric and Neuroradiology, Helios Hospital, Schwerin, Germany (H.-J.R.).

Objective: Respiratory motion can affect image quality and thus affect the diagnostic accuracy of CT images by masking or mimicking relevant lung pathologies. CT examinations are often performed during deep inspiration and breath-hold to achieve optimal image quality. However, this can be challenging for certain patient groups, such as children, the elderly, or sedated patients.

View Article and Find Full Text PDF

Objective: The aim of this single-center case series is to demonstrate that an ultra-low dose (ULD) can be routinely achieved in the hybrid operating room in standard endovascular aortic repair (EVAR) for infrarenal abdominal aortic aneurysm by adjusting the manufacturer's predefined imaging parameters, hardware configurations and user protocols (including benchmarking).

Methods: The hybrid operating room manufacturer predefined EVAR software setup of the dose exposure control software (OPTIQ, Siemens Healthineers, Forchheim, Germany) at our university medical center was screened for possible improvements regarding radiation dose application. Tests on a water-equivalent as well as polymethyl methacrylate phantom model to assess the impact of technical settings were performed, including comparison of settings for exposure control software, different magnification, collimation configurations and detector distance.

View Article and Find Full Text PDF

Objectives: This study evaluates the performance of a clinical dual-source photon-counting computed tomography (PCCT) system in quantifying iodine within calcified vessels, using 3D- printed phantoms with vascular-like structures lined with calcium.

Methods: Parameters assessed include lumen diameters (4, 6, 8, 10, and 12 mm), phantom sizes (S: 20×20 cm, M: 25×25 cm, L: 30×40 cm, XL: 40×50 cm, representing the 99th percentile of US patient sizes), and iodine concentrations (2, 5, and 10 mg/mL). Scans were performed at radiation dose levels of 5, 10, 15, and 20 mGy to systematically evaluate iodine quantification accuracy and spectral imaging performance.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!