Measurements were performed with various LiF based TLDs on board seven Polish aircraft, flying long-distance or middle-distance routes. All of the 7LiF detectors used (various types of 7LiF:Mg,Ti and 7LiF:Mg,Cu,P detectors), which measure the non-neutron component of the radiation field, produced consistent results. It was found that the characteristics of the TLD response (ratio of different detector responses, glow curve shapes) after doses of radiation at flying altitudes differ from those obtained after exposure at the CERN facility (CERF), suggesting a lower contribution of densely ionising radiation. The neutron induced TL signal was also more affected by the thickness of the holder, suggesting the presence of a softer neutron energy spectrum at flight altitudes. Further in-flight and CERF exposures of detectors are planned to resolve these issues.